Appendix A Specifications
Agility System
Operator’s Manual
145
Appendix A Specifications
A.1
System Specifications
Number of Plates
12 (additional packs can be added using
continuous load capacity)
Number of Sample Tubes
200
Number of Reagents
Up to 16 SmartKit
TM
Reagent Packs
simultaneously (additional packs can be
added using continuous load capacity)
Number of Pipettes
2 ( reagent)
A.2
Sample Pipetting Specifications
Sample Tip Size
300 µL
Sample Pipetting Volume
10-
300 µL
Time to Dispense
11
minutes (typical) 50 µL of 96 samples
from plate to sample tubes or deep well
plates
Sample Pipetting Precision
≤ 3% CV (at any operating volume above 10
µL)
Sample Pipetting Accuracy
± 2% of target volume
Dilution Range
1 part in 199 one-stage solution
1 part in 39,601 two-stage solution
Number of Sample Tips
20 racks of 112 tips (additional racks can be
added using continuous load capacity)
Sample Tube Dimensions
10-17 mm diameter external dimension
40-100 mm depth
A.3
Reagent Pipetting Specifications
Reagent Tip Size
12
00 µL
Reagent Pipetting Volume
20-
1200 µL
Number of Reagent Tips
1 rack of 98 tips (additional racks can be
added using continuous load capacity)
Reagent Pipetting Precision
≤ 3% CV at 10 shots at any volume in
operating range (single shot mode)
Reagent Pipetting Accuracy
± 2% of target volume (single shot mode)
Содержание Agility Automated ELISA System
Страница 6: ...Table of Contents 6 Agility System Operator s Manual This page intentionally left blank ...
Страница 8: ...About this Manual 8 Agility System Operator s Manual This page intentionally left blank ...
Страница 20: ...Chapter 1 Overview 20 Agility System Operator s Manual Figure 1 17 USB Hub Setup Figure 1 18 USB Hub Setup ...
Страница 24: ...Chapter 2 Safety 24 Agility System Operator s Manual This page intentionally left blank ...
Страница 42: ...Chapter 3 System Design 42 Agility System Operator s Manual This page intentionally left blank ...
Страница 52: ...Chapter 4 Operating the Agility System 52 Agility System Operator s Manual Figure 4 9 Wash Assay Screen ...
Страница 74: ...Chapter 4 Operating the Agility System 74 Agility System Operator s Manual Figure 4 28 Consumables Status ...
Страница 82: ...Chapter 4 Operating the Agility System 82 Agility System Operator s Manual This page intentionally left blank ...
Страница 98: ...Chapter 6 The Graphical User Interface 98 Agility System Operator s Manual Figure 6 1 The Administrator Screen ...
Страница 101: ...Chapter 6 The Graphical User Interface Agility System Operator s Manual 101 Figure 6 4 The Version Report Screen ...
Страница 144: ...Chapter 7Error Handling 144 Agility System Operator s Manual This page intentionally left blank ...
Страница 148: ...Appendix A Specifications 148 Agility System Operator s Manual This page intentionally left blank ...
Страница 150: ...Appendix B Consumables and Spare Parts 150 Agility System Operator s Manual This page intentionally left blank ...