No. 60025, Issue 5. January 1999
Diagnostics and Fault Finding 9
48
Wizard Diagnostics
The Programming Configuration Diagnostic (known as the Wizard) tool is used
to provide diagnostics for the Dolphin.
To View Diagnostics
1.
Enter the Wizard’s Main Menu screen as described in the Wizard
Installation Sheet / Online Help.
2.
Select the ‘Diagnostics’ menu.
The menu displays the following options:
Status Report
Print Status Report
Chair Log
Print Chair Log
Erase Chair Log
Print Chair Log
prints the Status Report followed by the Chair Log. These
reports should be sent along with a faulty controller to a Service Centre.
contact a Dynamic Sales and Service Centre (refer to section 14).
Status Report
The Status report gives you the current status of the wheelchair, including
faults and other warning conditions currently active. Pressing «?» Or selecting
‘Info’ will display further information about the condition. Some conditions in
the Status Report are not caused by actual faults, but are only temporary
conditions e.g. a motor lead was not connected when the DX System was
turned on and driving was attempted, producing a Motor Fault.
To View Status Report
1.
Perform steps 1. and 2. above.
2.
Select ‘Status Report’.
3.
From the ‘Modules Attached’ menu, select ‘UCM Remote’.