RMO25W Getting Started
6 Getting Started
6.1 Connecting a Test Object to RMO25W
Before you connect a test object to RMO25W, make sure that:
• the test object is disconnected or separated from its circuit in accordance with
the national safety regulations and is properly grounded to protective earth,
• RMO25W itself is properly grounded. To do so, connect the grounding screw
on the back of RMO25W’s to PE using a grounding cable.
• That one end of measured windings was properly grounded.
C A U T I O N
FOR SAFETY REASONS, ALWAYS ESTABLISH THE CONNECTION OF
DEVICE WITH TEST OBJECT IN THAT WAY AS AT THE FIGURE 6-1.
DO NOT DISCONNECT LEADS BEFORE LED DISCHARGE AND THE
BUZZER ARE OFF.
Figure 6-1:
Connecting a test
object to RMO25W
With RMO25W turned off,
connect RMO25W to the
test object in such a way
that the measuring cables
from the "Voltage Sense"
sockets are attached as
close as possible to R
X
,
and in between the
current feeding cables.
That way, resistance of
both cables and clamps
is almost completely
excluded from the
resistance measurement.
To take care about
polarity while
connecting measuring
cables.