14
RMO200D Getting Started
In the
Result
Menu, by pressing the
START
button the
Demagnetization
process will start. The whole
demagnetization process does not require modification of any connection, it is automatic and very quick.
After the Demagnetization is finis
hed, on the display appears the message „Demagnetization done“ for 3
seconds and the last results will be shown on the display again.
Now there are two choices:
Starting a new test with a different current value:
–
press the
STOP
button to change to the starting menu (
Single
,
Contin
or
CC
),
–
set a new current value and/or another memory position,
–
confirm your entries by pressing the
ENTER
button,
–
start the test by pressing the
START
button.
Starting a new test with the same current value:
–
press the
STOP
button to change to the starting menu (
Single
,
Contin
or
CC
),
–
press the
ENTER
button to change to the
Ready
state,
–
press the
START
button to run a test.
The current value remains the same as in the previous test, the memory position increases
by one.
If the
Rmax
menu is set to option
On
, once a Single or CC test is finished, RMO200D automatically
changes to the
Rmax
to display the test results in case the measured resistance is equal or greater than
maximum assigned value.
Figure 6-13:
The
Rmax
showing the test
results
Measured resistance of test object
Figure 6-11:
The
Result
menu
showing the test results
obtain in Continuous
test
Memory position that
holds the results
Max. test current reached
Voltage drop on the test
object
Figure 6-12:
The
Result
menu
showing the test results
obtain in CC test
Measured resistance of test object
Memory position that
holds the results
Max. test current reached
Voltage drop on the test
object
Содержание RMO200D