16
RMO200D Getting Started
6.8 Deleting Results of Previous Tests
In the
Memory
menu, press
ENTER
.
Press
ENTER
once more to delete all saved measurements from the memory positions.
Press
STOP
to
leave the
Memory
menu without deleting any measurements.
6.9 Duty Cycles
During tests, RMO200D generates a high DC current that heats up the test set. To prevent RMO200D from
overheating, certain duty cycles apply depending on the used test current.
S i n g l e T e s t
Test current
(A)
Cooling time between
first 3 tests (sec)
Cooling time after
4 tests (sec)
5, 10, 20, 50,
100, 150, 200
0
0
C o n t i n u o u s T e s t
Test current
(A)
Maximal test
duration
time (sec)
Cooling time
(sec)
5, 10, 20, 50,
100
300
0
150, 200
150
1 x test duration
A built-in control prevents these cooling times from being skipped. If you try to start a test within the cooling
period, the display shows "Wait" and a timer counting down. After the cooling period has elapsed, start the
test using the set test parameters by pressing the
START
button. Cooling of RMO200D is supported by a
built-in fan that is automatically activated every time a test is started from the
Ready
state. It continues
running 5 minutes after the test is finished.
Table 6-16:
Cooling time
between
single
tests
Figure 6-15: The
Memory
menu showing an option to
delete all saved
measurements from the
memory positions
Table 6-17:
Cooling time
between
continuous
tests