APT3700N Smart Pressure Transmitter for Nuclear Service Operation Manual M3700N-K01E
27
Step Recording Way
(1) Step Recording needs function :
12 Zero Adjustment, 22 Change Upper Range
Value, 23 Change Lower Range Value, 24 Damping Second functions
needs
functions of step recording.
(2) Step Recording Way : First select the Addition(
10
n
) and after changing the desired
value of Increase/decrease of changing value.
(3) For example of inserting 3810 : select 1000 increase -> increase 1000 3 times ->
select 100 increase -> increase 100 per 8 -> select 10 increase -> 10 times increase per
1
(4) Increase, Decrease Step selection : Message of SelInc is shown at the button of
LCD. Select Step with Zero Button : While pressing the Zero button, 10unit is changed.
After having a step setting, the span button should be pressed of performing 2 times.
(5) Zero, Span wants a value setting change : the bottom LCD has a VALUE message
shown When pressing the Zero button should increase the setting clause in an increase
Step When pressing the Span, the setting step decreases from the 1st clause After
hanging to a value wanted, Zero+Span button should be pressed to carry out the 1
st
clause.
(6) Repeat the forward number (4) and (5) of the desired value setting finally in the
value of desire value in the finished setting of fulfilling number(4)of the Zero+Span
button of finishing the input procedure.
Function Accompanying Performance
ZERO TRIM
- Press the ZERO+SPAN button and perform the pressing menu.
- When the
1 TRIM
message is shown, the SPAN button should be pressed to
move to the lower menu.
- If the
11 Z-TRIM
message is shown, the SPAN button is pressed to function
ZERO TRIM.
- Zero Adjustment : PV should be revised to 14
- ZERO+SPAN button is pressed to execute the menu
-
1 TRIM
message is shown, the span button is pressed to move to the lower
menu.
- If
11 Z-TRIM
message is shown, the Zero button is pressed to change menu.
- If
12 Z-ADJ
message is shown, press the San button to perform the Zero
Adjustment.
- If the
SelInc
message is shown, the expression of LCD
10.0
presses the Zero
button repeatedly and if the LCD
shows
10.0,
Span button is pressed to revise