Page 47
Changing the Input Fuse
The Terranova® 752A contains one slow-blow Type T, 2 A fuse used to interrupt AC power in
the event of overcurrent from the electrical outlet. The fuse is held in the fuse block located
within the AC power module, below the power cord socket. Recommended replacement fuses
are Bussman GDC-2A and Littelfuse 218 002.
To change the fuse:
1. Turn unit OFF via the
POWER
switch
2. Remove power cord and ion pump cable(s)
3. Locate the fuse block immediately below the power cord socket (see
4. Press the tab of the fuse block and withdraw the assembly
5. Inspect and replace faulty fuse
6. Reinsert fuse assembly into power module
7. Push fuse assembly into place until assembly tab "clicks"
When ready for use, user should re-connect the ion pump cable(s) and power cord. Thereafter,
Terranova® 752A operation may be resumed.
Figure 14.
Terranova® 752A AC power module
User should wait at least 15 seconds after turning OFF the
Terranova® 752A before removing the power cord
WARNING!