Page 42
Checksum Function
The checksum function is changed via the internal DIP Switch S3. Although DIP Switch S3 has eight
individual ON/OFF switches labeled 1 through 8, only switch S3-5 is utilized. Checksum function
is by default disabled. Checksum function can also be bypassed by sending a
00
checksum value
when enabled.
T
o change the checksum function:
1. Turn unit OFF via the
POWER
switch
2. Remove power cord and ion pump cable(s)
3. Remove the six #6-32 screws along the edge of the top cover
4. Carefully lift the top cover and disconnect the fan plug
5. Remove top cover
6. Locate DIP Switch S3 (see
7. Set DIP Switch S3-5 to desired function:
ON
: Checksum enabled
OFF
: Checksum disabled (default)
8. Connect the fan plug and replace top cover
When ready for use, user should re-connect the ion pump cable(s) and power cord. Thereafter,
the control unit may resume normal operation.
User should wait at least 15 seconds after turning OFF the
Terranova® 752A before removing the power cord
WARNING!
A checksum value is not required for communication
when checksum function is disabled.
NOTE