23
2013/10/21
9000-608-21/30
15.4 Reading out the image data
Description of the main control panel
ON/OFF button
Ready for operation
Data transmission
Fault
• Switch on the CR 43 at the ON/OFF button
• Start the computer
• Start the software
The sequence of the read-out pro-
cedure can differ, depending on the
software used See also the manual of
the software used
• Select the object type and recording mode
• Activate the CR 43 in the software
Observe the LEDs at the cassette slide
of the scanner
Green LED on the cassette slide is lit:
• Place the cassette into the cassette slide and
push into the appliance straight
The cassette is automatically locked and the
image plate drawn in for the scanning proc-
ess The cassette then moves back slightly,
but remains locked
This movement closes the light protec-
tion flaps
Yellow LED on the cassette slide is lit:
Cassette drawn in and locked Read-out proce-
dure running
The appliance must not be exposed
to vibrations during the read-out
procedure.
Green LED on the cassette slide flashes:
• Remove cassette from the cassette slide