Use
14. Cassettes
The scanner may only be operated
in combination with cassettes from
Dürr NDT. Other cassettes are not
compatible with the appliance and
could damage the appliance.
Cassettes
• protect the image plate from light, soiling and
mechanical damage
• prevent the erasure of image data on the im-
age plate
• are positioned at the cassette slide of the ap-
pliance
• Are designed for approx 25,000 cycles After
this time the cassettes should be replaced in
order to avoid incorrect diagnoses or defects
• can be briefly exposed to higher surface
temperatures as the surface is made of alu-
minium
14.1 Identification of the cassettes
General identification:
The size, order number
and identification number are shown on a label
on the narrow side
Active side:
The active side is not labelled
Passive side:
The passive side is marked
"
BACK SIDE
"
14.2 Handling of cassettes
Inspect cassettes for visible damage Dam-
aged cassettes (eg visible scratches that could
impair the diagnosis) must be replaced
• Do not bend or expose to extreme pressure
• Do not scratch or soil
• Do not immerse cassettes in liquids and do
not spray with liquids We recommend the use
of protective sleeves, Order No see point 41
6
7
EN
Содержание CR 43 NDT
Страница 1: ...CR 43 NDT HD CR 43 NDT Installation and Operating Instructions 9000 608 21 30 EN ...
Страница 2: ...2 ...
Страница 14: ...14 9000 608 21 30 2013 10 21 7 Functional layout 1 2 3 4 5 6 7 8 9 10 11 EN ...
Страница 33: ......
Страница 34: ...34 9000 608 21 30 2013 10 21 EN ...
Страница 35: ...35 2013 10 21 9000 608 21 30 EN ...