USER MANUAL
Micromotor for oral surgery
01_26/04/2018
Page 8 of 15
Pic. 7.
Function keys:
Key 1
Set in succession the parameter to modify (SPEED, TORQUE, PUMP, REDUCT): the adjustable
parameter is highlighted.
Key 2
Increase the highlighted parameter.
Key 3
Decrease the highlighted parameter.
Every adjustment is automatically saved in the current program, you don’t need to press any key for confirmation.
Even if a parameter is highlighted it’s possible to start the micromotor (by the foot-switch) or press any key.
Key 4
Selection of the programs (in sequence from 1 to 10): the current use program is shown on display.
Key 5
Led activation.
Switch on/off the led (when led is turned on the symbol is highlighted:
).
If pressed for 2 seconds key n.5 activates/deactivates “motor+led” function: this function
automatically switch on the led when the micromotor starts. On the display the led symbol is circled:
.
Key 6
Activation/deactivation “reverse” function (rotation way changing), when “reverse” is activated the
symbol is highlighted:
For reduction higher than 4:1 the “reverse” function is activable only if the set rotation speed
(SPEED) is lower or equal to 50 rpm.
Key 7
start/stop the peristaltic pump (when pump is activated the symbol is highlighted:
). If pressed
for 2 seconds key n.7 activates/deactivates “motor+pump” function: this function automatically start
the pump with the micromotor. On the display the pump symbol is circled:
.