USER MANUAL

Micromotor for oral surgery

01_26/04/2018

Page 5 of 15

INSTALLATION

• 

Take off the tape and open the box. Carefully take out the unit and accessories along with their original 
packaging and place them on a flat surface. Cut the protective material and remove the packaging for the unit 
and accessories. Leave enough space free around the unit to assure adequate ventilation.

• 

Check the equipment to assure it was not damaged before installing it. Damage incurred during shipment must 
be reported to your sales representative within 24 hours from when the goods were received. 

• 

After opening the packaging, make sure all the parts specified in the paragraph “DESCRIPTION OF VERSIONS 
AND COMPONENTS” are provided.

• 

The serial number stamped on the back of the unit must match the one written on the warranty card attached to 
this manual. 

When installing the device follow the safety rules given below:

• 

This equipment can be operated only by specially trained persons. Do not allow third parties to tamper with the 
controls. This instrument is not a toy. Do not play with it or use it for purposes other than those it is specifically 
designed for.

• 

Do not place the equipment near direct or indirect sources of heat.

• 

Do not put the equipment in humid environments or in contact with any type of liquids.

• 

Make sure enough air is able to circulate around the equipment to cool it down.

• 

It is risky and severely forbidden to alter or in any case modify the characteristics of the product. DSI Dental 
Solutions shall not be held responsible for any personal injury or equipment damage caused by failure to heed 
this precaution.

• 

Do not use the device near cellular phones.

Operating conditions:

• 

Temperature: 10° C /40° C.

• 

Humidity: 30% to 75%.

• 

Pressure: 500hPa to 1060hPa.

• 

After shipment allow the device to get accustomed to the climate in the work environment.

• 

The power supply may fluctuate not more than +/– 10% of the rated voltage.

• 

The micromotor cannot run continuously for more than 30 sec., with minimum 10 sec. intervals.

Transport and storage conditions:

• 

Temperature: –10° C /70° C.

• 

Humidity: 10% to 90%.

• 

Pressure: 500hPa to 1060hPa.

 

Before connecting the unit, make sure the voltage indicated on the plate found on the rear panel corresponds to 

the mains voltage. 

 

To avoid the risk of electrical shock this device must be connected only to an approved electrical system 

properly grounded. 

Содержание MINI SURGERY MOTOR

Страница 1: ...USER MANUAL Micromotor for oral surgery 01_26 04 2018 Page 1 of 15 USER MANUAL DSI MICROMOTOR FOR ORAL SURGERY 0476...

Страница 2: ...on the patients check the pre set operating parameters 8 Use only contra angles handpieces that conform with the essential requirements of European Directives and bear the CE mark as specified in the...

Страница 3: ...while the operator risks electrocution If liquids are split on the equipment proceed as follows Shut off the equipment and unplug the power cord Immediately dry the equipment casing outside with a rag...

Страница 4: ...1 Control unit 2 Micromotor with led 3 Variable speed pedal 4 Power supply cord 5 Micromotor support 6 Protective plug for sterilization 7 Bar for physiological solution Pic 2 Pic 3 1 Main switch 2 Po...

Страница 5: ...e it for purposes other than those it is specifically designed for Do not place the equipment near direct or indirect sources of heat Do not put the equipment in humid environments or in contact with...

Страница 6: ...rol unit Assembly the handpiece contra angle or straight on the micromotor and insert the micromotor into the support Use only handpieces with E Type connection ISO 3964 that comply as medical devices...

Страница 7: ...of control unit Connect the power supply cord to a grounded socket Switch on the device by the main switch USE All the adjustments are to be made with the motor stopped The display shows the followin...

Страница 8: ...lay Key 5 Led activation Switch on off the led when led is turned on the symbol is highlighted If pressed for 2 seconds key n 5 activates deactivates motor led function this function automatically swi...

Страница 9: ...ION AND STERILIZATION WARNINGS The operators performing the cleaning and sterilization operations must be adequately protected and trained Limitations to Repeated Processes Repetition of processes has...

Страница 10: ...special requirements Packaging Insert the protective cap on the micromotor The micromotor can be sterilized see instructions below only if provided with the protective cap accessory supplied List of...

Страница 11: ...ective The DSI Mini Surgery Motor device must function in the electro magnetic field specified below The customer user of DSI Mini Surgery Motor must guarantee that it is used in this environment DSI...

Страница 12: ...es continued operation during power mains interruption it is recommended that the product be powered from a separate power supply UPS etc Magnetic field at network frequency 50 Hz 3 A m 3 A m Magnetic...

Страница 13: ...s must be less than 3 V m DSI Mini Surgery Motor is intended for use in an electromagnetic environment in which radiated radiofrequency disturbances are controlled The DSI Mini Surgery Motor user and...

Страница 14: ...cycling treatment and compatible environmental disposal helps to avoid possible negative effects on the environment and health and favors the re use and or recycling of the materials it is composed of...

Страница 15: ...and ss mm ii Follow the instructions for use Attention carefully read the user manual before use It is afraid of humidity RAEE appliance separate collection object BF type electro medical device Fragi...

Отзывы: