USER MANUAL
Micromotor for oral surgery
01_26/04/2018
Page 5 of 15
INSTALLATION
•
Take off the tape and open the box. Carefully take out the unit and accessories along with their original
packaging and place them on a flat surface. Cut the protective material and remove the packaging for the unit
and accessories. Leave enough space free around the unit to assure adequate ventilation.
•
Check the equipment to assure it was not damaged before installing it. Damage incurred during shipment must
be reported to your sales representative within 24 hours from when the goods were received.
•
After opening the packaging, make sure all the parts specified in the paragraph “DESCRIPTION OF VERSIONS
AND COMPONENTS” are provided.
•
The serial number stamped on the back of the unit must match the one written on the warranty card attached to
this manual.
When installing the device follow the safety rules given below:
•
This equipment can be operated only by specially trained persons. Do not allow third parties to tamper with the
controls. This instrument is not a toy. Do not play with it or use it for purposes other than those it is specifically
designed for.
•
Do not place the equipment near direct or indirect sources of heat.
•
Do not put the equipment in humid environments or in contact with any type of liquids.
•
Make sure enough air is able to circulate around the equipment to cool it down.
•
It is risky and severely forbidden to alter or in any case modify the characteristics of the product. DSI Dental
Solutions shall not be held responsible for any personal injury or equipment damage caused by failure to heed
this precaution.
•
Do not use the device near cellular phones.
Operating conditions:
•
Temperature: 10° C /40° C.
•
Humidity: 30% to 75%.
•
Pressure: 500hPa to 1060hPa.
•
After shipment allow the device to get accustomed to the climate in the work environment.
•
The power supply may fluctuate not more than +/– 10% of the rated voltage.
•
The micromotor cannot run continuously for more than 30 sec., with minimum 10 sec. intervals.
Transport and storage conditions:
•
Temperature: –10° C /70° C.
•
Humidity: 10% to 90%.
•
Pressure: 500hPa to 1060hPa.
Before connecting the unit, make sure the voltage indicated on the plate found on the rear panel corresponds to
the mains voltage.
To avoid the risk of electrical shock this device must be connected only to an approved electrical system
properly grounded.