USER MANUAL
Micromotor for oral surgery
01_26/04/2018
Page 2 of 15
INTENDED USE
The DSI Mini Surgery Motor product are brushless micromotor for oral surgery. The use of DSI Mini Surgery Motor is
allowed only to specialized personnel (licensed medical specialists) in hospitals, nursing homes, clinics and private
practices.
SAFETY RULES
1. Clean, disinfect and sterilize all the sterilizable parts (ref. par. Maintenance) after each treatment as instructed in
this manual.
2. All new or repaired products are in not-sterilized conditions: before using on the patient the sterilizable parts
must be disinfected and sterilized.
3. Check periodically the condition of the supply cord. Replace it with a genuine spare part when worn down.
4. Check periodically the condition of the micromotor handpiece cord. Replace it with a genuine spare part when
worn down.
5. Connect to the device only the accessories/components (supplied from DSI Dental Solutions company) that
are part of the original product (see par. Components). Using accessories/components not indicated or put in
the market by DSI Dental Solutions as spare parts for replacement may cause an increase of electromagnetic
emissions or a decrease of the device’s electromagnetic immunity.
6. Do not use the instrument where strong electromagnetic fields are present (cellular phones, receiver-transmitter
radios, microwaves, etc…).
7. In the event of malfunctioning, shut off the apparatus and contact the manufacturer’s Customer Service
Department. Before using the equipment on the patients, check the pre-set operating parameters.
8. Use only contra-angles/handpieces that conform with the essential requirements of European Directives and
bear the CE mark as specified in the Medical Devices Directive 93/42/EEC.
9. Only qualified personnel can open the control unit case.
10. Do not use the equipment if the case is damaged.
11. To disconnect electrically the equipment from the main disconnect the power cord and/or press the main switch.
DSI Dental Solutions guarantees to provide on request the technical documentations about components parts
lists, schemes of electrical connections, information about range, accuracy and precision of the displayed values.
This documentation can be useful to skilled personnel in order to repair the equipment’s part that the manufacturer
considers repairable.
SAFETY INFORMATION
• Before attempting to operate the device, read and become familiar with the contents of the User’s Manual supplied
with the instrument.
• Failure to follow these recommendations, even partly, may cause equipment damage along with malfunctions.
Following the instructions and recommendations given by the manufacturer is the key to top performance and
efficient technical support, when necessary.
• This manual must be kept nearby the instrument so that it can be rapidly consulted whenever necessary. This
manual must accompany the instrument whenever it changes hands or is re-sold so that the new owner can
familiarize him/herself about the safety rules and precautions and how to properly operate the instrument.
• These rules and precautions are furnished for safety reasons. They must be carefully read in their entirety before
attempting to install and use the instrument.