USER MANUAL

Micromotor for oral surgery

01_26/04/2018

Page 2 of 15

INTENDED USE

The DSI Mini Surgery Motor product are brushless micromotor for oral surgery. The use of DSI Mini Surgery Motor is 
allowed only to specialized personnel (licensed medical specialists) in hospitals, nursing homes, clinics and private 
practices.

SAFETY RULES

1.  Clean, disinfect and sterilize all the sterilizable parts (ref. par. Maintenance) after each treatment as instructed in 

this manual.

2.  All new or repaired products are in not-sterilized conditions: before using on the patient the sterilizable parts 

must be disinfected and sterilized.

3.  Check periodically the condition of the supply cord. Replace it with a genuine spare part when worn down.
4.  Check periodically the condition of the micromotor handpiece cord. Replace it with a genuine spare part when 

worn down.

5.  Connect to the device only the accessories/components (supplied from DSI Dental Solutions company) that 

are part of the original product (see par. Components). Using accessories/components not indicated or put in 
the market by DSI Dental Solutions as spare parts for replacement may cause an increase of electromagnetic 
emissions or a decrease of the device’s electromagnetic immunity.

6.  Do not use the instrument where strong electromagnetic fields are present (cellular phones, receiver-transmitter 

radios, microwaves, etc…).

7.  In the event of malfunctioning, shut off the apparatus and contact the manufacturer’s Customer Service 

Department. Before using the equipment on the patients, check the pre-set operating parameters.

8.  Use only contra-angles/handpieces that conform with the essential requirements of European Directives and 

bear the CE mark as specified in the Medical Devices Directive 93/42/EEC.

9.  Only qualified personnel can open the control unit case.
10. Do not use the equipment if the case is damaged.
11. To disconnect electrically the equipment from the main disconnect the power cord and/or press the main switch.

 

DSI Dental Solutions guarantees to provide on request the technical documentations about components parts 

lists, schemes of electrical connections, information about range, accuracy and precision of the displayed values. 
This documentation can be  useful to skilled personnel in order to repair the equipment’s part that the manufacturer 
considers repairable.

SAFETY INFORMATION

• Before attempting to operate the device, read and become familiar with the contents of the User’s Manual supplied 
with the instrument.
• Failure to follow these recommendations, even partly, may cause equipment damage along with malfunctions. 
Following the instructions and recommendations given by the manufacturer is the key to top performance and 
efficient technical support, when necessary.
• This manual must be kept nearby the instrument so that it can be rapidly consulted whenever necessary. This 
manual must accompany the instrument whenever it changes hands or is re-sold so that the new owner can 
familiarize him/herself about the safety rules and precautions and how to properly operate the instrument.
• These rules and precautions are furnished for safety reasons. They must be carefully read in their entirety before 
attempting to install and use the instrument.

Содержание MINI SURGERY MOTOR

Страница 1: ...USER MANUAL Micromotor for oral surgery 01_26 04 2018 Page 1 of 15 USER MANUAL DSI MICROMOTOR FOR ORAL SURGERY 0476...

Страница 2: ...on the patients check the pre set operating parameters 8 Use only contra angles handpieces that conform with the essential requirements of European Directives and bear the CE mark as specified in the...

Страница 3: ...while the operator risks electrocution If liquids are split on the equipment proceed as follows Shut off the equipment and unplug the power cord Immediately dry the equipment casing outside with a rag...

Страница 4: ...1 Control unit 2 Micromotor with led 3 Variable speed pedal 4 Power supply cord 5 Micromotor support 6 Protective plug for sterilization 7 Bar for physiological solution Pic 2 Pic 3 1 Main switch 2 Po...

Страница 5: ...e it for purposes other than those it is specifically designed for Do not place the equipment near direct or indirect sources of heat Do not put the equipment in humid environments or in contact with...

Страница 6: ...rol unit Assembly the handpiece contra angle or straight on the micromotor and insert the micromotor into the support Use only handpieces with E Type connection ISO 3964 that comply as medical devices...

Страница 7: ...of control unit Connect the power supply cord to a grounded socket Switch on the device by the main switch USE All the adjustments are to be made with the motor stopped The display shows the followin...

Страница 8: ...lay Key 5 Led activation Switch on off the led when led is turned on the symbol is highlighted If pressed for 2 seconds key n 5 activates deactivates motor led function this function automatically swi...

Страница 9: ...ION AND STERILIZATION WARNINGS The operators performing the cleaning and sterilization operations must be adequately protected and trained Limitations to Repeated Processes Repetition of processes has...

Страница 10: ...special requirements Packaging Insert the protective cap on the micromotor The micromotor can be sterilized see instructions below only if provided with the protective cap accessory supplied List of...

Страница 11: ...ective The DSI Mini Surgery Motor device must function in the electro magnetic field specified below The customer user of DSI Mini Surgery Motor must guarantee that it is used in this environment DSI...

Страница 12: ...es continued operation during power mains interruption it is recommended that the product be powered from a separate power supply UPS etc Magnetic field at network frequency 50 Hz 3 A m 3 A m Magnetic...

Страница 13: ...s must be less than 3 V m DSI Mini Surgery Motor is intended for use in an electromagnetic environment in which radiated radiofrequency disturbances are controlled The DSI Mini Surgery Motor user and...

Страница 14: ...cycling treatment and compatible environmental disposal helps to avoid possible negative effects on the environment and health and favors the re use and or recycling of the materials it is composed of...

Страница 15: ...and ss mm ii Follow the instructions for use Attention carefully read the user manual before use It is afraid of humidity RAEE appliance separate collection object BF type electro medical device Fragi...