USER MANUAL
Micromotor for oral surgery
01_26/04/2018
Page 10 of 15
Preparation before cleaning
Switch off the device using the switch and unplug the power cord.
Disconnect all components from the control unit and separate all accessories:
• micromotor.
• micromotor support.
• pedal.
Cleaning
Clean the surface of all components and accessories with a clean, soft cloth moistened
with a cleaning solution (pH 6-9).
Automated cleaning is not recommended.
Disinfection
If you intend to disinfect a component, use a non-aggressive disinfectant solution with a
neutral pH (pH7), following the instructions provided by the manufacturer of the solution;
dry with a clean, non-abrasive cloth.
Pay particular attention to the drying of the electrical connectors of the micromotor.
Drying
Not applicable.
Maintenance
No special requirements.
Inspection and function tests
No special requirements.
Packaging
Insert the protective cap on the micromotor.
The micromotor can be sterilized (see instructions below) only if provided with the
protective cap (accessory supplied).
List of sterilisable components:
• micromotor with protective cap.
• micromotor support.
Seal all individually sterilizable accessories in a disposable sterilization bag.
Sterilization
The class B steam autoclaving process is recommended according to EN 13060.
Type of cycle: 3 times Pre-vacuum (minimum pressure 60 mBar).
Minimum sterilization temperature: 132° C (range 0° C ± 3° C).
Maximum sterilization temperature: 135° C.
Minimum sterilization time: 4 minutes.
Minimum drying time: 20 minutes.
All sterilization steps must be performed by the operator in accordance with EN ISO
17665-01.
Wait until the sterilized components, especially the handpiece and the micromotor,
cool completely before using them.
Storage
Store sterilized materials in a dry and dust-free place.
Additional Information
No additional information.
Manufacturer contact
Ref. last page of this user manual.