Preparation before cleaning

 

Switch off the device using the switch and unplug the power cord.

Disconnect all components from the control unit and separate all accessories:
•  micromotor.
•  micromotor support.
•  pedal.

Cleaning

Clean the surface of all components and accessories with a clean, soft cloth moistened 
with a cleaning solution (pH 6-9).

 

Automated cleaning is not recommended.

Disinfection

If you intend to disinfect a component, use a non-aggressive disinfectant solution with a 
neutral pH (pH7), following the instructions provided by the manufacturer of the solution; 
dry with a clean, non-abrasive cloth.

 

 

Pay particular attention to the drying of the electrical connectors of the micromotor.

Drying

Not applicable.

Maintenance

No special requirements.

Inspection and function tests

No special requirements.

Packaging

Insert the protective cap on the micromotor.

 

The micromotor can be sterilized (see instructions below) only if provided with the 

protective cap (accessory supplied).
List of sterilisable components:
•  micromotor with protective cap.
•  micromotor support.

Seal all individually sterilizable accessories in a disposable sterilization bag.

Sterilization

The class B steam autoclaving process is recommended according to EN 13060.

Type of cycle: 3 times Pre-vacuum (minimum pressure 60 mBar).
Minimum sterilization temperature: 132° C (range 0° C ± 3° C).
Maximum sterilization temperature: 135° C.
Minimum sterilization time: 4 minutes.
Minimum drying time: 20 minutes.
All sterilization steps must be performed by the operator in accordance with EN ISO 
17665-01.

 

 

Wait until the sterilized components, especially the handpiece and the micromotor, 

cool completely before using them.

Storage

Store sterilized materials in a dry and dust-free place.

Additional Information

No additional information.

Manufacturer contact

Ref. last page of this user manual.

Содержание MINI SURGERY MOTOR

Страница 1: ...USER MANUAL Micromotor for oral surgery 01_26 04 2018 Page 1 of 15 USER MANUAL DSI MICROMOTOR FOR ORAL SURGERY 0476...

Страница 2: ...on the patients check the pre set operating parameters 8 Use only contra angles handpieces that conform with the essential requirements of European Directives and bear the CE mark as specified in the...

Страница 3: ...while the operator risks electrocution If liquids are split on the equipment proceed as follows Shut off the equipment and unplug the power cord Immediately dry the equipment casing outside with a rag...

Страница 4: ...1 Control unit 2 Micromotor with led 3 Variable speed pedal 4 Power supply cord 5 Micromotor support 6 Protective plug for sterilization 7 Bar for physiological solution Pic 2 Pic 3 1 Main switch 2 Po...

Страница 5: ...e it for purposes other than those it is specifically designed for Do not place the equipment near direct or indirect sources of heat Do not put the equipment in humid environments or in contact with...

Страница 6: ...rol unit Assembly the handpiece contra angle or straight on the micromotor and insert the micromotor into the support Use only handpieces with E Type connection ISO 3964 that comply as medical devices...

Страница 7: ...of control unit Connect the power supply cord to a grounded socket Switch on the device by the main switch USE All the adjustments are to be made with the motor stopped The display shows the followin...

Страница 8: ...lay Key 5 Led activation Switch on off the led when led is turned on the symbol is highlighted If pressed for 2 seconds key n 5 activates deactivates motor led function this function automatically swi...

Страница 9: ...ION AND STERILIZATION WARNINGS The operators performing the cleaning and sterilization operations must be adequately protected and trained Limitations to Repeated Processes Repetition of processes has...

Страница 10: ...special requirements Packaging Insert the protective cap on the micromotor The micromotor can be sterilized see instructions below only if provided with the protective cap accessory supplied List of...

Страница 11: ...ective The DSI Mini Surgery Motor device must function in the electro magnetic field specified below The customer user of DSI Mini Surgery Motor must guarantee that it is used in this environment DSI...

Страница 12: ...es continued operation during power mains interruption it is recommended that the product be powered from a separate power supply UPS etc Magnetic field at network frequency 50 Hz 3 A m 3 A m Magnetic...

Страница 13: ...s must be less than 3 V m DSI Mini Surgery Motor is intended for use in an electromagnetic environment in which radiated radiofrequency disturbances are controlled The DSI Mini Surgery Motor user and...

Страница 14: ...cycling treatment and compatible environmental disposal helps to avoid possible negative effects on the environment and health and favors the re use and or recycling of the materials it is composed of...

Страница 15: ...and ss mm ii Follow the instructions for use Attention carefully read the user manual before use It is afraid of humidity RAEE appliance separate collection object BF type electro medical device Fragi...

Отзывы: