Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These specifications may not be applicable in all situations. Electromagnetic propagation is affected by absorption and 
reflection from structures, objects and persons.
(a): The electromagnetic field strengths of fixed radiofrequency emitters, such as base stations for mobile telephones (cellular / 
cordless), mobile radios, amateur radio, AM/FM radio broadcasts and TV broadcasts cannot be determined exactly by theory. To 
assess the electromagnetic environment due to fixed radiofrequency emitters, an electromagnetic environment measurement 
must be made. If the measured radiofrequency field strength in the immediate environment where the product is used exceeds 
the compliance level specified above, the performance of the product must be tested to verify whether it conforms to the 
specification. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating 
the product.
(b): In the 150 kHz to 80 MHz frequency range, the electromagnetic field strengths must be less than 3 V/m.

DSI Mini Surgery Motor is intended for use in an electromagnetic environment in which radiated radiofrequency
disturbances are controlled.
The DSI Mini Surgery Motor user and/or installer can help prevent electromagnetic interference by maintaining a 
minimum distance between portable and mobile radiofrequency communications equipment emitters and the DSI 
Mini Surgery Motor, according to the maximum output power of the equipment, as recommended in the table below.

Rated max. power of the emitter (W)

SEPARATION DISTANCE IN METRES (M) ACCORDING TO EMITTER FREQUENCY

150 kHz to 80 MHz

80 MHz to 800 MHz

800 MHz to 2.5 GHz

0.01

0.12

0.12

0.23

0.1

0.37

0.37

0.74

1

1.17

1.17

2.33

10

3.69

3.69

7.38

100

11.67

11.67

23.33

For emitters rated at max. power not listed above, the recommended separation distance d in meteres (m) can be estimated 
using the equation applicable to the frequency of the emitter, where P is the max. power rating of the emitter in watts (W) 
according the manufacturer.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies. 
Note 2: These specifications may not be applicable in all situations. Electromagnetic propagation is affected by absorption and 
reflection from structures, objects and persons.

Cable lengths:
•  Power supply cord  – 2.0 m.
•  Footswitch cord  – 1.8 m.
•  Micromotor cord  – 2.0 m.

 

Содержание MINI SURGERY MOTOR

Страница 1: ...USER MANUAL Micromotor for oral surgery 01_26 04 2018 Page 1 of 15 USER MANUAL DSI MICROMOTOR FOR ORAL SURGERY 0476...

Страница 2: ...on the patients check the pre set operating parameters 8 Use only contra angles handpieces that conform with the essential requirements of European Directives and bear the CE mark as specified in the...

Страница 3: ...while the operator risks electrocution If liquids are split on the equipment proceed as follows Shut off the equipment and unplug the power cord Immediately dry the equipment casing outside with a rag...

Страница 4: ...1 Control unit 2 Micromotor with led 3 Variable speed pedal 4 Power supply cord 5 Micromotor support 6 Protective plug for sterilization 7 Bar for physiological solution Pic 2 Pic 3 1 Main switch 2 Po...

Страница 5: ...e it for purposes other than those it is specifically designed for Do not place the equipment near direct or indirect sources of heat Do not put the equipment in humid environments or in contact with...

Страница 6: ...rol unit Assembly the handpiece contra angle or straight on the micromotor and insert the micromotor into the support Use only handpieces with E Type connection ISO 3964 that comply as medical devices...

Страница 7: ...of control unit Connect the power supply cord to a grounded socket Switch on the device by the main switch USE All the adjustments are to be made with the motor stopped The display shows the followin...

Страница 8: ...lay Key 5 Led activation Switch on off the led when led is turned on the symbol is highlighted If pressed for 2 seconds key n 5 activates deactivates motor led function this function automatically swi...

Страница 9: ...ION AND STERILIZATION WARNINGS The operators performing the cleaning and sterilization operations must be adequately protected and trained Limitations to Repeated Processes Repetition of processes has...

Страница 10: ...special requirements Packaging Insert the protective cap on the micromotor The micromotor can be sterilized see instructions below only if provided with the protective cap accessory supplied List of...

Страница 11: ...ective The DSI Mini Surgery Motor device must function in the electro magnetic field specified below The customer user of DSI Mini Surgery Motor must guarantee that it is used in this environment DSI...

Страница 12: ...es continued operation during power mains interruption it is recommended that the product be powered from a separate power supply UPS etc Magnetic field at network frequency 50 Hz 3 A m 3 A m Magnetic...

Страница 13: ...s must be less than 3 V m DSI Mini Surgery Motor is intended for use in an electromagnetic environment in which radiated radiofrequency disturbances are controlled The DSI Mini Surgery Motor user and...

Страница 14: ...cycling treatment and compatible environmental disposal helps to avoid possible negative effects on the environment and health and favors the re use and or recycling of the materials it is composed of...

Страница 15: ...and ss mm ii Follow the instructions for use Attention carefully read the user manual before use It is afraid of humidity RAEE appliance separate collection object BF type electro medical device Fragi...

Отзывы: