Calibration Procedures
Note
: GE can provide a calibration service that is traceable
to international standards.
GE recommends returning the instrument to the
manufacturer or an approved service agent for calibration.
If using an alternative calibration facility, check that it uses
the standards that follow.
Use only original parts supplied by the manufacturer.
10.1
Before Starting
To perform an accurate calibration:
•
Use the calibration equipment specified in
Table 10-1
.
•
Calibrate in a stable temperature environment: 21 ± 1°C
(70 ± 2°F).
Note:
Before starting a calibration procedure, leave the
equipment in the calibration environment for a
minimum period of two hours.