User Manual
Druck DPI620G-IS
22 Feb 2017
116M5464 Revision A
Page 104 of 248
Sequence to Upload and Download File
Step
Procedure
1. Connect Standard USB lead (as supplied) to the Druck.
DPI620G-IS calibrator.
2. Connect the lead to the USB port on the computer that
has the calibrator manager installed.
3. Use 4 Sight / Intecal v10 to set-up the procedure and
create a work order for the device.
The procedure includes the parameters for the
calibration, the number of test points, the relationship
and pass/fail tolerance.
4. Use the Download button in calibrator manager to
Download the file to the Druck DPI620G-IS calibrator. A
communications symbol will be displayed at the bottom
of the screen.
5. Select
DASHBOARD
>>
DOCUMENTING
>>
RUN
PROCEDURE
6. In the Results window, select the filename specified in
calibrator manager.
7. Enter the User ID and the DUT Serial Number.
8. Tap on the Start button . The Procedure sets up the
necessary Channel options e.g. mA and Volts.
9. Use the Take Reading button at each point specified by
the procedure. A prompt shows for each point.
10. When all the readings are complete, tap on the Exit
Button . Look at the results on the display (As
found/As Left).
11. To complete the process, use the calibrator manager to
Upload the file back into the 4 Sight / Intecal v10
database