Electrical warnings
•
To prevent electrical shocks or damage to the instrument,
do not connect more than 30V CAT I between the
terminals, or between the terminals and the ground
(earth). Any connection must be compliant with the
terminal input/output parameters.
•
External circuits should have appropriate insulation to the
mains.
•
This instrument uses a Lithium-Ion (Li-Ion) battery pack.
To prevent an explosion or fire, do not short circuit, do not
disassemble, and keep safe from damage.
•
To prevent an explosion or fire, use only the GE specified
battery (IO620G-IS-BATTERY), power supply (Part: IO620-
PSU) and battery charger (Part: IO620G-IS-CHARGER).
•
To prevent battery leakage or heat generation, only use:
-
the battery charger (IO620G-IS-CHARGER ) and
power supply IO620-PSU in the temperature
range 0°C to 40°C (32°C to 104°F).
•
The power supply IO620-PSU input range is 100 – 240Vac
+/-10%, 50 to 60Hz, 250mA, installation category CAT II.
•
Note that the operating and storage temperature range
of the power supply (IO620-PSU) does not match that of
the DPI620G-IS battery charger. (IO620-PSU operating
temperature range 0°C to +40°C, storage temperature
range -40°C to +70°C).
•
Position the power supply so as not to obstruct the supply
disconnecting device.