2.
Some methods need values to be entered. Use the
alpha/numeric keypads.
3.
A drop down menu can be used for method selection
options.
4.
Some methods require input from the DPI620G-IS
instrument channels. A drop down menu displays the
channels as follows:
•
CH1
•
CH2
•
P1
•
P2
•
IDOS
5.
On completion the procedure will return to the HART®
application. If required use the Abort button.
Method Example - Self-test
1. To confirm that the transmitter is functioning
correctly, navigate to the Test device folder.
2. Select the Test device folder.
3. Select OK.
The self-test executes.