Method Example - Analog Trim
The Druck DPI620G-IS can perform an analogue trim on
the 4 to 20 mA loop without connecting to any external
reference meters.
1.
Navigate to the calibration folder
Figure 6-11 Calibration Folder
2.
Select Trim analog method.
3.
Follow the on-screen instructions. If channel 2 is set to
Measure Current (15V), this can be used to supply the
reference meter value.