Manual, Ultra High Sensitivity Aerosol Spectrometer (UHSAS)
DOC-0210 Rev E-4
4 8
© 2017 DROPLET MEASUREMENT TECHNOLOGIES
12.0
Configuration Tab
This tab displays the instrument configuration parameters. Note that some of these parameters
are protected and cannot be changed unless the user logs in by typing a password in the
Password
field and then clicking on the
Enter Password
button. To revert to normal operation
(i.e., not a password-protected mode), click on the
Revert Mode
button. Figure 29 shows the
Configuration
tab in Calibration mode.
Clicking on
Revert to Factory Configuration
reverts all the parameters to their original default
values. These parameters are stored in separate configuration file,
factory.ini
, which
should not be modified.
The
Configuration File Date
field lists the date the configuration file was last modified in
YYYYMMDD format.
Clicking the
Save
button saves the present instrument configuration to a file. When this button
is clicked, the system will prompt the user for a file name. The file must be saved in the directory
C:\DMT\Spectrometer Support.
As stated above, the factory provided configuration
file,
factory.ini
, should not be overwritten.
Figure 29: Configuration Tab