Use the handles, which are located about halfway down each side of the case, to lift the
instrument from the case, and place it on a stable working area. The front legs of the
instrument are hinged so that the front of the instrument case can be elevated (see
Warning:
The edges of the instrument legs are sharp. Handle carefully to avoid
puncturing cables, and always ship the instrument with the legs folded to avoid damage
to the shipping case.
2)
The instrument will be shipped with 1.5 inches of extra tubing on the inlet line. This
tubing makes a “tail” that fits over the sample needle. See Figure 9.
If the sample tubing needs to be reinstalled, use 1/8” (~3.2 mm) outside diameter (OD)
conductive silicon or urethane tubing.
3)
Connect the keyboard/track-pad to a USB port on the front panel. The USB ports are
located just below the floppy/CD combo drive.
4)
Make sure the instrument ON/OFF switch is in the off position. The ON/OFF switch is a
rocker-type toggle located on the back of the instrument near the power plug.
5)
Connect the power cable to the wall power outlet. Input power to the instrument can
range from 100-240 VAC 50/60Hz.
6)
Connect the UHSAS instrument to an Ethernet network, if desired. The connector (RJ-45
female) for that cable is located on the back of the instrument.
7)
If desired, use the serial port connection (9 pin D-connector) located on the back of the
instrument to connect to an external data system.
Figure 9: Extra Tubing on Inlet Line