background image

deutsch

english

25

fr

ançais

24

PCU LED N

2

24

25

The PCU LED N

2

 (Post Curing Unit LED) 

is a light polymerization device for the 

curing of components which have been 

created by means of 3D printing. The 

following device properties are used in 

the curing process:

a) LED technology

b)  Electronic control  

(micro-controller)

c)  Logging and monitoring of the  

process parameters

d) Simple control concept

e) Process reliability

f)  Different curing environments  

(air, vacuum, nitrogen)

g) Open system

Device properties:

a) LED technology

The advantages of the LED techno-

logy are that a permanent light of 

one wavelength is emitted. The per-

manently emitted light radiation can 

be captured by a standard UV sensor 

and thus allows for the measurement 

and the observation of the light out-

put in the exposure chamber. Further-

more, LED lamps do not produce any 

heat which could cause the material 

properties of the plastic component to 

be cured to be negatively influenced. 

Another advantage of the LED technol-

ogy is that the stated operating life of 

the LED lamps is approx. 20,000 hours 

and is accompanied by low operating 

and maintenance costs.

b) Electronic control (micro-controller)

The use of a micro-controller allows for 

f) Curing environment

The components to be cured can be 

cured under different environments. 

The environment to be set depends 

on the application and can be found 

in table 1.

1. Device description

the control and reading of the current, 

pressure and temperature sensors.

c) Logging and Monitoring

The information which is recorded by 

the current, pressure and temperature 

sensors represents the used process 

parameters (e.g. light on/off; vacuum 

on/off; nitrogen on/off). These process  

parameters are recorded, saved and 

transferred to a USB storage medium. 

This allows for the transparent trace-

ability of the production chain of the 

medical device.

d) Simple control concept

The device is operated by the remov-

able rotary button. The corresponding 

option is selected by turning the but-

ton. The selected option is operated by 

pressing. Longer pressing opens the 

parent and sub-folder menu. Another 

removable element is the handle which 

is used to open the lid of the exposure 

chamber. Both elements can be cleaned 

with the appropriate cleaning agents 

after removal.

e) Process safety

To be able to guarantee a high pro-

cess liability, the control software is  

subdivided into two levels. Level 1 

is accessible to the user and process 

owner. Level 2 is only accessible to the 

process owner. In level 1 only a prede-

fined process can be selected. In level 2, 

for example, a process can be defined 

and an option can be switched on and/

or off.

g) Open system

In the PCU LED N

2

, 10 freely program-

mable storage spaces can be used.

Table 1: Curing environment for different uses

Environment

Use

PCU LED N

2

 nitrogen

Vacuum

Curing of laboratory 

products

Nitrogen

Curing of  

medical devices

Weight of the overall package 

11.1 kg

Weight of the device 

9.3 kg

Dimensions of the overall package  H 220 x W 455 x D 370 mm

Dimensions of the device 

H 110 x W 389 x D 276 mm

Polymerization chamber 

H 65 x W 150 x D 150 mm

Voltage 

100–240 V

Nominal frequency 

50–60 Hz

2. Technical data

Содержание PCU LED N2

Страница 1: ...hnology Dreve Firmengruppe Max Planck Stra e 31 59423 Unna Germany deutsch english Lichtpolymerisationsger t f r den CAD CAM 3D Druck Prozess Light polymerization device for the CAD CAM 3D printing pr...

Страница 2: ...mation on the use of the protective gas 29 7 Functional elements 30 8 Operation 31 9 Maintenance 37 10 Monitoring 38 11 Troubleshooting plan 38 12 Exploded view 40 13 Spare parts list 41 14 General in...

Страница 3: ...s Mikrocontrol lers k nnen Strom Druck und Tem peratursensoren angesteuert und aus gelesen werden c Logging und Monitoring Die Informationen die von den Strom Druck und Temperatursensoren auf gezeichn...

Страница 4: ...rom 16 A je Leiter IEC 61000 3 2 2014 Deutsche Fassung EN 61000 3 2 2014 EN 61000 3 3 2013 Elektromagnetische Vertr glichkeit EMV Teil 3 3 Grenzwerte Begren zung von Spannungs nderungen Spannungsschwa...

Страница 5: ...23 Das Ger t nicht betreiben wenn das Geh use des Ger tes nicht geschlossen und verschraubt ist 24 Eigenm chtige Umbauten und Ver nderungen sind aus Sicherheitsgr nden unzul ssig Stellen Sie vor der I...

Страница 6: ...lt in zwei Ebenen Die erste Ebene ist die Anwenderebene in der der Bediener verschiedene Optionen ausw hlen und best tigen kann Die zweite Ebene ist die Prozesseignerebene in dieser kann der Prozessei...

Страница 7: ...n Parametersatz Statusfenster Durch nochmaliges Best tigen wird der Belichtungsprozess gestartet 1 Programmname 2 Returntaste 3 Belichtungsdauer 4 Druck mbar 5 Belichtungsicon ON OFF 6 Belichtungsatmo...

Страница 8: ...f r das Anwenderlevel Einstellen der Prozessparame ter Dr cken des Drehknopfes 5 Sekunden 8 4 Bedienung Prozessebene Einstellen der Prozessparameter 1 Returntaste 2 Programmname frei definierbar durch...

Страница 9: ...des Geh uses mit einem milden Reinigungsmittel w chentlich Wartungseinheit Wartungsh ufigkeit berpr fen ob die LED Panels gewechselt werden m ssen w chentlich mit Programm LED Test berpr fen ob die L...

Страница 10: ...ziehen Abschaltung nach 150 Sekunden Deckeldichtung der Belichtungskammer undicht Erneuern der Deckeldichtung Schl uche des Vakuum systems undicht Erneuern der Schl uche Vakuumpumpe defekt Austauschen...

Страница 11: ...33 51050 Steckverschraubung Legris 34 51079 3M BUMPON KLEBEFU SJ 5023 35 51283 Cordeset NEMA Netzstecker5 15 USA gerade ohne Abbildung 36 521092 Glasscheibe Nachh rteeinsatz PCU LED 37 521091 Rahmen N...

Страница 12: ...n durch von ihr erm chtigte Personen ausge f hrt werden Da wir stetig an der Weiterentwicklung unserer Erzeugnisse arbeiten behalten wir uns technische nderungen vor Die Herstellergarantie betr gt 2 J...

Страница 13: ...gging and Monitoring The information which is recorded by the current pressure and temperature sensors represents the used process parameters e g light on off vacuum on off nitrogen on off These proce...

Страница 14: ...3 3 2013 Electromagnetic compatibility EMC Part 3 3 Limits Limitation of voltage changes voltage fluctuations and flicker in public low voltage supply systems for equipment with rated current 16 A per...

Страница 15: ...ribed operating and main tenance conditions specified in these operating instructions must be com plied with Please observe the general accident prevention regulations when working with the PCU LED N2...

Страница 16: ...form different device configurations 1 Switching on the power plug at the back of the device brings you directly to the user level 2 Switching on the power switch at the back of the device and the si...

Страница 17: ...eter set in the status window Status window Additional confirmation starts the exposure process 1 Programme name 2 Return button 3 Exposure time 4 Pressure mbar 5 Exposure icon ON OFF 6 Exposure atmos...

Страница 18: ...t activated for the user level Setting the process para meters press the rotary button 5 seconds 8 4 Operation Process level Setting the process parameters 1 Return button 2 Programme name freely defi...

Страница 19: ...connection mains connection daily Cleaning of the device with a mild detergent weekly Maintenance unit Maintenance frequency Check whether the LED panels must be or changed weekly with program LED Tes...

Страница 20: ...tch LED defective Exchange of the LED 04 No nitrogen flow Nitrogen valve at the gas cylinder not opened Warning The device is only o be operated with a maximum pressure of 5 bar 70 psi Open nitrogen v...

Страница 21: ...521091 Rahmen Nachh rteeinsatz PCU LED 6 D C B A E F G H E A B C D F 5 4 3 2 1 7 8 9 10 11 12 1 2 3 4 5 6 7 8 DRV13078 Dreve ProDiMed GmbH Max Planck Str 31 D 59423 Unna Germany Tel Nr 02303 88 07 0 w...

Страница 22: ...he transfer of risk due to faulty or unintended use exces sive loading unsuitable equipment and any such chemical electrochemical or electrical influences which are not pro vided in accordance with th...

Страница 23: ...un microcontr leur l on peut exciter et lire les capteurs de courant de pression et de temp rature c Journalisation et suivi des para m tres de processus Les informations enregistr es par les capteurs...

Страница 24: ...limites des courants harmoniques cou rant appel par les appareils 16 A 3 D claration de conformit CE EN ISO 12100 2010 rectification 1 de 12100 2011 03 DIN EN 60204 1 2006 A1 2009 rectification 1 zu D...

Страница 25: ...Les Transformations et modifica tions arbitraires sont interdites pour des raisons de s curit Avant la mise en service assurez vous qu il n y a pas de dommages dus au transport V rifiez si le bo tier...

Страница 26: ...vis en deux niveau le premier niveau est le niveau de l utilisateur o l op rateur peut s lectionner et confirmer diff rentes options Le second niveau est le niveau du propri taire des processus dans l...

Страница 27: ...a m tres s lectionn Fen tre d tat Le processus d exposition est lanc par une confirmation r it r e 1 Nom du programme 2 Touche Retour 3 Dur e d exposition 4 Pression mbar 5 Symbole d exposition MARCHE...

Страница 28: ...non activ pour le niveau de l utilisateur R glage des param tres de processus pression du bou ton tournant 5 secondes 8 4 Commande niveau des processus R glage des param tres de processus 1 Touche Ret...

Страница 29: ...o tier avec un d tergent doux toutes les semaines Unit de maintenance Fr quence de mainte nance V rifier si les panneaux de LED doivent tre remplac s toutes les semaines avec le programme test LED V r...

Страница 30: ...e de la chambre d exposi tion fuit Renouveler le joint d tanch it du couvercle Tuyaux du syst me du vide fuient Renouveler les tuyaux Pompe vide d fec tueuse Remplacer la pompe vide 03 Couvercle ouver...

Страница 31: ...50937 Drosselventil 6 01 18 1N G1 8 31 51012 Dichtring aus Perbunan 32 51030 MECANYL ROHR RW LT 2A0031 S 33 51050 Steckverschraubung Legris 34 51079 3M BUMPON KLEBEFU SJ 5023 35 51283 Cordeset NEMA Ne...

Страница 32: ...ons sont r alis s par des per sonnes autoris es par elles Dans la mesure o nous d veloppons nos produits en permanence nous nous r servons le droit d y apporter des modi fications techniques La garant...

Страница 33: ...Dreve Design Rev 221119 D221119 200 1854 Dreve Firmengruppe Max Planck Stra e 31 59423 Unna Germany www dreve com...

Отзывы: