English
Device-specific components
The device-specific components must be removed
from the device and, if necessary, disassembled.
Removing the CO
2
absorber
1
Remove the CO
2
absorber:
– CLIC absorber (disposable)
Or
– Reusable CO
2
absorber
CLIC absorber (disposable absorber)
Follow the instructions for use for the CLIC
absorber.
1
Press the release button (A). The CLIC adapter
flips open.
2
Pull the CLIC absorber (C) upwards out of the
holder (B).
3
Flip the holder back until it clicks into place.
Reusable absorber
1
Unscrew the CO
2
absorber (A) from the
breathing system.
2
Remove and dispose of the optional disposable
dust filter (B).
3
Empty the used soda lime. Dispose of
according to the instructions for use for the soda
lime.
4
Remove the absorber insert (C) from the
absorber container. The sealing ring (D)
remains on the absorber insert.
5
Prepare the CO
2
absorber for cleaning and
disinfection in the washer-disinfector.
068
A
B
C
018
013
A
B
D
C