English
16
Supplement Zeus
Infinity Empowered
Machine cleaning with thermal disinfection
Use a washer-disinfector that meets the
requirements of the standard ISO 15883. Dräger
recommends the use of a load carrier for
anesthesia accessories and ventilation
accessories. Follow the manufacturer's instructions
for the washer-disinfector.
Components:
– Breathing system
– Housing (upper part and lower part)
– Expiratory port, inspiratory port
– APL valve
– Additional air valve
– Absorber container
– Absorber insert
– TurboVent 2 blower module
– Flexible breathing bag arm (optional)
Prerequisites:
– The washer-disinfector has been prepared in
accordance with the manufacturer's
instructions.
Step
Agent
Manufacturer Concentration
Temperature
Contact time
Cleaning
Neodisher
MediClean
Forte
Dr. Weigert
According to the
manufacturer's
instructions
Min. 55 °C
(131 °F)
Min. 10 min
Disinfecting
–
–
–
Min. 90 °C
(194 °F)
Min. 5 min
Drying
–
–
–
–
Drying time de-
pends on the
load