Classifications for reprocessing
Classification of medical devices
The classification depends on the intended use of
the medical device. The risk of infection
transmission through the application of the product
to the patient without proper reprocessing is the
basis of the Spaulding classification.
Classification of device-specific components
Follow the instructions for use for the components.
The following classification is a recommendation
from Dräger.
Non-critical
– Device surface including screen
– Trolley and holders
– Removable parts of the main device
– Breathing system mount
Semi-critical
– Breathing system
– Housing (upper part and lower part)
– Expiratory port, inspiratory port
– APL valve
– Additional air valve
– Flexible breathing bag arm (option)
– TurboVent 2 blower module
– Absorber container and absorber insert
Critical
The device does not contain any components
classified as critical.
Classification
Explanation
Non-critical
Components that come into contact only with skin that is intact
Semi-critical
Components that carry breathing gas or come into contact with mucous mem-
branes or pathologically altered skin
Critical
Components that penetrate skin or mucous membranes or come into contact with
blood