Supplement Zeus
Infinity Empowered
Before reprocessing
Observe before disassembly
1
Switch off the device and all devices connected
to it.
2
Disconnect all power plugs.
Patient-specific accessories and consumables
The patient-specific accessories and consumables
must be removed from the device and, if necessary,
disassembled.
Reusable products:
If the reusable product has its own instructions
for use, perform reprocessing in accordance
with the separate instructions for use.
If no separate instructions for use are available
for the reusable product, perform reprocessing
in accordance with the instructions in this
supplement.
Disposable products:
Dispose of the disposable products.
Removing patient-specific accessories and
consumables
Remove the following accessories:
– Breathing hoses
– Breathing bag
– Filters
– Water traps
Removing the filter
Remove the filter (A) from the nozzle on the Y-
piece, inspiratory port, or expiratory port.
Removing the breathing bag
Remove the breathing bag and hose (B).
046
A
C
C
B