![Dräger VITALERT 2000 Скачать руководство пользователя страница 47](http://html.mh-extra.com/html/drager/vitalert-2000/vitalert-2000_operators-instruction-manual_2526846047.webp)
APPENDIX 5: VITALINK COMMANDS
The Vitalert 2000 acts on the following
Vitalink commands:
UPDATE REQUEST COMMANDS
Code Command
24H
Request current data
25H
Request current low alarm limits
26H
Request current high alarm limits
27H
Request current active alarms
28H
Request current date and time
CONTROL COMMANDS
Code Command
31H
Enable audio alarm annunciation
32H
Disable audio alarm annunciation
33H
Enable alarms
34H
Disable alarms
36H
Log data
38H
Clear trend data
39H
Invoke power-on default settings
51H
Initialize communications
CONFIGURATION COMMANDS
Code Command
20H
Request data configuration
21H
Request low alarm limit
configuration
22H
Request high alarm limit
configuration
23H
Request alarm configuration
The Vitalert 2000 issues the following
Vitalink commands:
UPDATE REQUEST COMMANDS
Code Command
24H
Request current data
27H
Request current active alarms
28H
Request current time and date
29H
Request available waveform
2AH
Request current waveform data
2BH
Stop waveform
CONTROL COMMANDS
Code Command
51H
Initialize communications