![Dräger VITALERT 2000 Скачать руководство пользователя страница 40](http://html.mh-extra.com/html/drager/vitalert-2000/vitalert-2000_operators-instruction-manual_2526846040.webp)
APPENDIX 2: VITALERT 2000 ALARM MESSAGES
CLASS
ALARM
MESSAGE
CONDITION
AUDIO
ALARM
CODE
ALARM
PRIORITY
W
HRT RATE LOW
Heart rate less than low
limit
CONT
05H
31
C
HRT RATE HI
Heart rate greater than
high limit
INT
21H
21
C
ART SYS LOW
Arterial systolic less than
low limit
INT
5FH
21
C
CVP MEAN LOW
CVP mean less than low
limit
INT
7DH
21
C
PA DIAS LOW
PA diastolic less than low
limit
INT
7FH
21
C
P2 SYS LOW
P2 systolic less than
lowlimit
INT
87H
20
C
ART SYS HI
Arterial systolic greater
than high limit
INT
69H
20
C
CVP MEAN HI
CVP mean greater than
high limit
INT
80H
20
C
PA DIAS HIGH
PA diastolic greater than
high limit
INT
82H
20
C
P2 SYS HIGH
P2 Systolic greater than
high limit
INT
88H
20
C
ECG PWR FAIL
No AC power and battery
low
INT
9AH
11
A
ECG AC FAIL
AC power failure/
disconnect
SINGLE
TONE
90H
7
A
ECG LL DISC
ECG left leg lead
disconnect
NONE
4FH
2
A
ECG RL DISC
ECG right leg lead
disconnect
NONE
8CH
2
A
ECG LA DISC
ECG left arm lead
disconnect
NONE
50H
2