OPERATING INSTRUCTIONS
Turning on the VITALERT 2000
The ON and STANDBY keys control the
power to the VITALERT 2000 (Fig. 11).
Pressing
the
ON
key
activates
the
VITALERT
2000,
which
responds
by
performing a series of self-diagnostic checks
to verify its ability to function properly. The
results of these tests are posted on the
screen.
At
the
conclusion
of
the
self-diagnostic tests, a summary of any
problems and the operational status are
posted on the display screen.
The operational status of the VITALERT
2000 may be any one of the following:
FUNCTIONAL
- The VITALERT 2000
initiates normal monitoring operation.
CONDITIONALLY FUNCTIONAL
- Any
minor deficiencies are posted, and the
operator must press the CONTINUE soft
key. A North American Dräger service
representative should be notified.
NON-FUNCTIONAL
- The self-diagnostics
reveal a problem in the VITALERT 2000. A
summary of the diagnostic results is posted
and normal operation is inhibited. A North
American Dräger service representative
should be notified immediately.
Each time the VITALERT 2000 is turned on,
it resets all alarm limits to their power-on
default values and disables the pacemaker
mark function, the Auto Print function and
the Auto Plot function. The VITALERT 2000
also selects the monitoring bandwidth, sets
the waveform speed to 25 mm/sec, and the
ECG scale to 10 mm/mV on power-up. All
other configuration settings are retained
from the last VITALERT 2000 configuration.
Refer to the previous chapter to make
changes
to
the
VITALERT
2000
configuration.
Keyboard Operation
A description of each key’s function is given
below.
ON
-
activates
the
VITALERT
2000
Monitor. The VITALERT 2000 can also be
remotely activated through its external
communications port. When interfaced to a
NARKOMED 3 anesthesia system, the
VITALERT 2000 is automatically activated
by the System Power switch.
STANDBY
- disables the VITALERT 2000.
The battery charger of the VITALERT 2000
continues
to
operate
as
long
as
the
VITALERT 2000 is plugged into an active
AC outlet.
ALARMS
ON/OFF
-
disables
all
VITALERT 2000 patient alarms (Heart Rate
and Blood Pressure). While alarms are
disabled, an advisory message is placed on
the VITALERT 2000 display screen. Alarms
can be re-enabled by pressing the ALARMS
DISABLE key again.
HELP
- invokes the Help mode to assist the
operator in understanding the operation of
the VITALERT 2000.
CONFIG
- invokes the Configuration mode
to allow the operator to customize the
operation of the VITALERT 2000.
FREEZE
- stops and saves a portion of the
current ECG waveform on the VITALERT
2000
Monitor
screen.
Real
time
data
continues to be accumulated and displayed
above the frozen ECG waveform.
SOFT KEYS
- used to manipulate the
Configuration, Help and Freeze screens.
LEAD
- used to select the desired limb lead
to be displayed (See
ECG Lead
section for
more details).
16