![Dräger VITALERT 2000 Скачать руководство пользователя страница 43](http://html.mh-extra.com/html/drager/vitalert-2000/vitalert-2000_operators-instruction-manual_2526846043.webp)
APPENDIX 4: TROUBLESHOOTING GUIDE
PROBLEM
POSSIBLE CAUSE
REMEDY
No ECG trace.
Cable disconnect.
Secure connections.
Lead disconnected from
electrode or cable.
No pressure trace.
Transducer improperly attached
to monitor.
Secure attachment.
Pressure channel not zeroed.
Zero the channel.
Unable to zero
pressure.
Faulty transducer.
Replace the transducer.
Pressure offset too high.
Verify that the catheter tip is as
close as possible to the same
elevation as the transducer
dome.
Verify that the transducer is in
the position of use and attached
to the fluid system.
Check the transducer
manufacturer’s instructions for
use.
Waveform greater than 5 mm
Hg, peak to peak.
Check stopcocks. Prior to
zeroing, be sure the pressure
transducer has been positioned
and connected for at least 5
minutes for stabilization.
Noisy ECG traces.
Loose or dry electrodes.
Apply new electrodes.
Defective electrode wires.
Replace the wires.
Patient cables or leads routed
too close to other electrical
devices.
Reroute other cables.
Electrosurgical
interference.
Inadequate skin prep prior to
electrode application.
Repeat skin prep and electrode
placement procedures.
Wrong cable type.
Replace the cable.
Cable is damaged.
35
Содержание VITALERT 2000
Страница 2: ...RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 4: ...RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 52: ...RETURN TO CD ROM TABLE OF CONTENTS ...
Страница 53: ...RETURN TO CD ROM TABLE OF CONTENTS ...