CONFIGURATION (continued)
VITALERT 2000 Monitoring Configuration
The Monitoring Options screen (Fig. 5)
controls how information is presented on the
Monitor screen.
Flashing Heart
The first option, Flashing Heart, activates a
flashing heart on the monitor screen as a
visual representation of R-wave detection.
P2 Label
The P2 Label option selects one of three
labels for the P2 pressure channel (P2 CVP,
P2 PA, or P2 ART).
Trace Speed
Trace Speed selects the sweep rate of the
waveforms. The sweep rate can be configured
for either 12.5 mm/sec, 25 mm/sec, or
50 mm/sec. On power-up, the sweep rate
defaults to 25 mm/sec.
P1 Scale
The P1 Scale option configures the P1
SCALE
key
for
manual
or
automatic
pressure waveform scale selection. See the
Operating
Instructions
section
of
this
manual for more information.
P2 Scale
The P2 Scale option configures the P2
SCALE
key
for
manual
or
automatic
pressure waveform scale selection. See the
Operating
Instructions
section
of
this
manual for more information.
Temperature Units
Selecting Temperature Units allows the
operator
to
display
the
measured
temperature in either Fahrenheit or Celsius.
HI
LO
P1
0
P2
SCALE
ALARMS
OFF
BLOOD PRESURE ALARMS
HEART RATE ALARMS
HI
LO
P1
0
SCALE
P2
FREEZE
OP27005
LEAD
SCALE
BASELINE AUTOSCALE
CONFIG
HELP
68
100
70
0
SELECT
MONITOR
ENTER
CVP
OFF
PA
25
50
AUTO
AUTO
MAN UAL
MAN UAL
OFF
MONITOR
DIAG
C
ART
F
O N
FLASHING HEART:
P2 LABEL:
TRACE SPEED:
P1 SCALE:
P2 SCALE:
TEMPERATURE UNITS:
PACEMAKER MARK:
BANDWIDTH:
12.5
CONFIGURE VITALERT 2000 MONITORING OPTIONS
O N
FIGURE
FIGURE 5:
5: MONITORING OPTIONS CONFIGURATION SCREEN
10