Mach LED 500DF
Dr. Mach
Lamps and Engineering
59190002
Edition 05
15.09.2014 / Bak We page 4/26
1.
Safety instructions
Pay attention to the instructions for use when handling the lamp.
WARNING:
This device has not been designed for use in potentially explosive areas.
According to the Medical Device Regulation the light is classified under class I.
Store the OT-light in its package for at least 24 hours in the respective room before mounting, in
order to equal temperature differences.
Please read the instructions for use carefully to make the most of your lighting system and to avoid
any damages to the device.
The lights may only be repaired and special assembly work may only be carried out on the
reflector or sockets by ourselves or a company that has been expressly authorised by us.
The manufacturer can only be made responsible for the safety of the light if repairs and alterations
are carried out by the manufacturer himself or a company that guarantees to observe the safety
regulations.
No modification of the lamp is allowed!
The manufacturer cannot be made liable for personal or material damages if the light is operated
inexpediently or incorrectly or used for purposes other than those for which it is intended.
The light is to be dismantled from the spring arm in reverse order to its assembly. This may only
be carried out after the spring arm has been adjusted in height at horizontal position since the
arm is under spring tension and can bounce up.
Make sure that the light is in perfect working order before every use.
Attention, external power supply!
The light works only with an external power supply 250VA or 200VA.
The external power supply used with the OT-light must be tested and validated according to
IEC 60601-1.
Attention!
A
main control switch must be installed for turning the system power-off. The switch must meet the
requirements of the standard IEC 61058-1 regarding rated voltage peaks of 4kV.
During the mounting of the OT-lights the entire system
(incl. the ceiling attachment) must be disconnected from mains!
A later dismounting of the lights from the spring arms or dismounting the sliding contacts inside the
arms is to be done ONLY AFTER DISCONNECTING THE ENTIRE SYSTEM FROM MAINS.
Otherwise the main control board will be damaged!