Mach LED 500DF
Dr. Mach
Lamps and Engineering
59190002
Edition 05
15.09.2014 / Bak We page
16/26
Inspection / durability
The sterilisable handle sleeve must be disposed
after 1000 sterilisation cycles or at the latest after
2 years and replaced with a new one.
The year of manufacture can be determined with
the help of a stamping on the inner side of the
handle sleeve (like shown in the photo). The
stamping in the photo shows the number 12,
which stands for the year 2012.
3.2 Lamp housing, protective disk
and support system
The OT-lamp MACH LED 500DF has a high-quality
surface, which can be cleaned with conventional
cleaning agents.
The protective disks
2
are made of a high-quality
plastic. Pay attention to the following during clean-
ing:
-
Never wipe over the protective disks
2
with a dry
cloth (always clean with a wet cloth).
Wipe the protective disks
2
after cleaning with
an antistatic, non-fluffy cloth.
2