DT803X

7

 The meter shows a voltage value > 1 V 

 Never perform the meas

-

urement of current if open circuit voltage to earth exceeds 600 V.

AC/DC amperage measurement

  1. Press and hold   button 

.

  2. Connect the red testing cable to 

10A

 socket 

, and the black cable to 

COM

 socket 

.

  3. Nacisnąć 3 razy przycisk 

SELECT

 

.

  4. Open the circuit, in which amperage will be measured and attach gauging 

cables to the gauged circuit.

  5. Read amperage value from the display 

Range

Resolution

Accuracy

Max

Frequency: 

40-1000 Hz

DC 1 A

1 mA

±1.0% + 5 digits

10 A

DC 10 A

10 mA

AC 1 A

1 mA

±1.2% + 6 digits

AC 10 A

10 mA

 Amperage measurements > 2 A - measurement time < 3 s (min. 15 

min interval).

Resistance measurement 

  1. Press and hold   button 

.

  2. Hold down the 

SELECT

 

 button for min. 2 s to turn on the auto range 

function. The display will show "AUTO". The meter will automatically adjust 

the type of measurement to resistance.

  3. Connect the red testing cable to  socket 

 

, and the black cable to 

COM

 socket 

.

  4. Attach gauging cables to resistor that will be measured.

  5. Read current’s amperage from the display 

 range

Resolution

Accuracy

Max

600 

0.1 

±1.3%+5 digits

60 M

6 k

60 k

10 

±1.0%+5 digits

600 k

100 

6 M

1 k

60 M

10 k

±1.5%+5 digits

 Prior to resistance measurement the circuit’s power supply must be 

switched off and all capacitors discharged.

Circuit continuity test

  1. Press and hold   button 

.

  2. Hold down the 

SELECT

 

 button for min. 2 s to turn on the auto range 

function. The display will show "AUTO". The meter will automatically adjust 

the type of measurement to the circuit continuity test.

  3. Connect the red testing cable to  socket 

 

, and the black cable to 

COM

 socket 

.

  4. Attach gauging cables to capacitor paying attention to polarization. 

 Continuity of the circuit shall be flagged with an audio signal when 

resistance is < 50 

.

Capacitance measurement

  1. Press and hold   button 

.

  2. Press the 

SELECT

 button 

 twice to select the nF function.

  3. Connect the red testing cable to  socket 

 

, and the black cable to 

COM

 socket 

.

  4. Attach gauging cables to capacitor paying attention to polarization.

  5. Read capacity value on display 

Range

Resolution

Accuracy

Max

6 nF

1 pF

±3.0%+5 digits

60 mF

60 nF

10 pF

600 nF

100 pF

F

1 nF

60 

F

1 0nF

600 

F

100 nF

±3.5%+5 digits

6 mF

F

±5.0%+6 digits

60 mF

10 

F

±10.0%+8 digits

 The tested capacitor should be discharged before the testing pro

-

cedure. 

Voltage detector NCV

  1. Press and hold   button 

.

  2. Press the 

SELECT

 button 

 to select the EF function. 

  3. When the voltage is detected in the range of AC 30-1000 V, a sound will be 

emitted and the display will show dashes "- - - -"

Flashlight / Display backlight 

Press and hold the 

 button 

 to switch the flashlight 



 and back

-

light 

 on or off.

Battery replacement

Replace the battery with a new one when the low battery symbol 

 

 

appears on the LCD display. Low battery level can cause erroneous meas

-

urements.

  1. Remove the back cover of the multimeter.

  2. Place new 2 x AAA batteries, paying attention to proper polarization.

 Before removing the back cover, disconnect all test leads from the 

meter and the measured circuit and the switch and the meter by 

holding down the 

 

 button 

 It is recommended to use batter

-

ies and fuses consistent with device’s specification . Used batter

-

ies as hazardous waste should be disposed of in a specially marked 

container in or sent to a selective collection point 

 Do not throw 

batteries into a rubbish bin 

 Do not use new and used batteries at 

the same time. 

Warranty

The warranty terms are available at http://www.dpm.eu/gwarancja

Allgemeines 

Multimeter mit automatischer Bereichswahl und berührungsloser Spannungs

-

detektor (NCV) 

 Messungen: AC/DC Bereich und Spannung, Widerstand, Kon

-

desatore 

 Dioden Test 

 Stromkreisdurchgangstest 

 true RMS Messung 

 

LCD-Display (4 Ziffern) 

 Hintergrundbeleu Taschenlampe 

 automa-

tische Abschaltung (15 min) 

 Anzeige der leeren Batterie 

  berspannungs

-

chutz 

 Versorgung: 2 x AAA Batterie.

Содержание DT803X

Страница 1: ...yczny miernik cyfrowy Auto range multimeter Multimeter mit automatischer Bereichswahl Automatinis nuotolio multimetras Automatick rozsahov multimetr Autom tiskais diapazona multimetrs Automaatne ulatu...

Страница 2: ...COM SELECT CAT III 600V 10sec MAX MAX 600V NCV EF OHM 0 1 60M 0 1 50 Hz 10Hz 20MHz 1 600V 0 60mF 100mA 10A NCV Hz...

Страница 3: ...AUTO MINMAX hFE m Mk Hz C F AV OUT nF...

Страница 4: ...zi pora eniem pr dem elektrycznym 14 Nie nale y u ywa urz dzenia gdy na wy wietlaczu pojawia si symbol wy czerpanej baterii Niski poziom baterii mo e powodowa b dne wskazania pomiar w 15 Przed wymian...

Страница 5: ...P a czarny prze w d gniazda COM O 4 Przy o y przewody pomiarowe do urz dzenia Ci g o obwodu zostanie zasygnalizowana sygna em d wi kowym przy oporno ci 50 Pomiar pojemno ci kondensator w 1 Nacisn i pr...

Страница 6: ...to faulty gauging indications 15 Make sure that the device is switched off before batteries are replaced 16 In case device is not used for a longer period of time the battery should be removed in orde...

Страница 7: ...utton 2 Press the SELECT button K twice to select the nF function 3 Connect the red testing cable to socket P and the black cable to COM socket O 4 Attach gauging cables to capacitor paying attention...

Страница 8: ...der Reichweite von Kindern 18 Das Ger t ist f r den Einsatz in Innenr umen bei Raumtemperatur bestimmt 19 Nutzen Sie das Ger t nicht direkt nach der Verlegung aus einem Raum mit hoher Luftfeuchtigkeit...

Страница 9: ...gezeigt werden wenn der Widerstand 50 ist Kapazit t von Kondensatoren Messung 1 Halten Sie die Taste gedr ckt 2 Dr cken Sie die SELECT Taste zweimal K 3 Stecken Sie das rote Messkabel in die Buchse P...

Страница 10: ...73 34 0 40o C 80 10o 50o 85 115 CATI ATI CATII CATII CATIII CATIII CATIV O ON OFF K SELECT L LCD 4 M HOLD N 10A O COM P Q R S HOLD NCV AC DC 1 2 SELECTK 2 3 P a COMO 4 5 L Max 40 1000 Hz DC 6 V 0 001...

Страница 11: ...1 k 60 M 10 k 1 5 5 1 2 SELECTK 2 3 P a COMO 4 50 1 1 2 SELECTK nF 3 P a COMO 4 5 L Max 6 nF 1 pF 3 0 5 60 mF 60 nF 10 pF 600 nF 100 pF 6 F 1 nF 60 F 1 0nF 600 F 100 nF 3 5 5 6 mF 1 F 5 0 6 60 mF 10...

Страница 12: ...18 renginys skirtas naudoti patalpose kambario temperat roje 19 Nenaudokite prietaiso i kart po jo atne imo patalp kurioje yra didel dr gm ema ar auk ta temperat ra Technin specifikacija Matavim tiksl...

Страница 13: ...inga po lis kuma 5 Perskaityti talpyklos kondensatoriaus verte disple juje L Diapazonas Rezoliucija Tikslumas Max 6 nF 1 pF 3 0 5 skaitmen 60 mF 60 nF 10 pF 600 nF 100 pF 6 F 1 nF 60 F 1 0nF 600 F 100...

Страница 14: ...ploty 23o C 5o C a pro vlhkost vzduchu 75 P esnost hodnoty indikace po et nejm n vy znamny ch slic 4 m stn LCD displej nap jen 2x AAA po et ode t 2 s pojistka F 10 A 600 V bzu k rozm ry 120 x 60 x 35...

Страница 15: ...ho krytu odpojte v echny testovac vodi e od m i e a m en ho obvodu a sp na e a m i e podr en m tla tka Doporu uje se pou vat baterie a pojistky v souladu se spe cifikac za zen Pou it baterie jako neb...

Страница 16: ...notas zemsprieguma instal cij m t d m k m jsaimniec b biroj vai darbn c s izmantojam s ier ces CAT III kategorie m en CAT III definuje bezpe nostn po adavky pro m en prov d n v obvodech instalac uvnit...

Страница 17: ...polu izvieto jumam Pirms aizmugur j v ci a no em anas atvienojiet visus testa vadus no skait t ja un m r t s des sl d a un skait t ja turot nospiestu pogu Ieteicams izmantot ier ces specifik cij m at...

Страница 18: ...aismojumu N 10A ligzda m r anas ligzda kas j pievieno ar sarkano kabe a amp ros m r jumu O COM SOCKET m r anas ligzda kas j pievieno ar melno kabeli P ligzda m r anas ligzda kas j pievieno ar sarkano...

Страница 19: ...nne testimisprotseduuri t hjaks laa dida Kontaktivaba pingedetektoriga NCV 1 Vajutage ja hoidke all nuppu 2 Vajutage nuppu SELECT K et valida EF funktsioon 3 Kad spriegums tiek noteikts diapazon no AC...

Страница 20: ...kykit s statym ir panaudot elektronini prietais nei meskite kartu su buitin mis atliekomis Tinkamas panaudot produkt r iavimas padeda suma inti j neigiam poveik aplinkai ir moni sveikatai Seznamte se...

Отзывы: