Dover WILDEN Saniflo VC4 Скачать руководство пользователя страница 10

S e c t i o n   5

T R o u b L E s H o o T i n g

thick  products  or  high  pumping  volumes  may  require 
you  to  increase  the  discharge  pressure .   the  discharge 
pressure will control the velocity of the product coming 
out of the pump and should be adjusted according to the 
desired  product  flow .   the  time  value  on  the  controller 
is  the  interval  that  the  pump  is  in  the  discharge  mode .  
A pump set to a short time interval will discharge small 
amounts  of  product  and  cycle  too  quickly .   A  pump  set 
to a long time interval will force too much product out of 
the tube and air lock the check valves .  Please note that on 
very fragile products like tenders or fillets, a longer stroke 
is more advantageous because the check valves will come 
in contact with fewer pieces of product .

•    Please note that in certain applications the HPVC pump 

may have to be primed to start the pump .  Applications 
that  may  require  priming  include  breast,  thigh  meat, 
tenders and other extremely thick products .  the pump 
can  be  primed  by  placing  either  your  hand  or  other 
object  (i .e .  plastic  bag)  over  the  exit  end  of  the  pipe 
connected  to  the  pump .    Once  the  pump  cycles  twice 
nothing else will be required .

trouble shooting guide

In  the  unlikely  event  that  your  pump  is  not  performing, 
please  see  the  chart  below  for  help  with  any  problems 
you might incur .

1. Problem

pump is not pulling when the plc indicates suction.

Cause

 Outlet  check  valve  may  be  lodged  open  with 

1. 

product causing the pump to suck air from piping .

 suction regulator is set too low .  (recommended: 

1. 

3 .8-5 bar (55-75 PsI) .

Low or no air pressure .

2. 

Air valves not functioning properly .

3. 

Dirty or defective vacuum generator .

4. 

Loose connection on piping .

5. 

Check valves misaligned .

6. 

Solution

 manually discharge the pump for 2 seconds to clear 

1. 

any  blockage .    If  this  does  not  help,  disassemble 
the pump and clear the check valves .

turn up suction regulator to 5 bar (75 PsI) .

2. 

Check to ensure that the air supply is adequate

3. 

 Disassemble  and  clean  vacuum  generator,  or 

4. 

replace unit

 ensure  that  all  tubing  or  plumbing  connections, 

5. 

both inside the control panel and at the pump, are 

seated properly and no air is leaking .

 Align check valve(s) hinge at top of tee . (refer to 

6. 

sec, 1-2)

trouble shooting guide

2. Problem

pump  is  in  suction  mode  and  will  not  cycle  to 
discharge mode.

Cause

PLC failure

1. 

Probe connection is bad .

2. 

Outlet check valve lodged open .

3. 

Probe set point is too high

4. 

Solution

 there are no user serviceable parts inside the PLC .  

1. 

replacement is recommended if it is found that the 
PLC has failed

 Inspect probe connections at both ends of wiring 

2. 

and repair if needed .

Purge pump .

3. 

Calculate probe set point and adjust PLC

4. 

trouble shooting guide

3. Problem

pump  stays  in  discharge  mode  and  will  not  cycle 
back to suction mode.

Cause 

 Probe may have product build up that is interfering 

1. 

with normal operation .

Pump wiring problem .

2. 

PLC failure

3. 

Probe set point is too low

4. 

Solution

 remove sensor cap from pump and clean probe to 

1. 

remove any product build up .

 Check all wiring, ensuring that there are no loose 

2. 

connections or corrosion

 there are no user serviceable parts inside the PLC .  

3. 

replacement is recommended if it is found that the 
PLC has failed

Calculate probe set point and adjust PLC 

4. 

S e c t i o n   6

A s s E M b L y

8

WILDEN PUMP & ENGINEERING CO., LLC

WIL-12120-E-01

Содержание WILDEN Saniflo VC4

Страница 1: ...R e f i n e y o u r p r o c e s s Engineering Opera ti o n Maint e n a n ce Saniflo Series METAL Pumps VC4 VC6 VC8 WIL 12120 E 01 ...

Страница 2: ...ribution System 3 Section 4 Dimensional Drawings 4 Section 5 Suggested INSTALLATION OPERATION TROUBLESHOOTING 5 Section 6 Assembly 9 Section 7 Exploded View Parts Listing sANIFLOtm VC4 12 sANIFLOtm VC6 14 sANIFLOTM VC8 16 VC Pump controller 18 Section 8 vc pump controller schematic 20 ta b le o f c o n te n t s ...

Страница 3: ... CAUTION The Saniflo VC pump is not submersible Mount control panel where it is not directly in the wash down area CAUTION Always wear safety glasses when operat ing pump If rupture of wetted components occurs material being pumped may be forced from pump and may create a hazardous situation CAUTION Before any maintenance or repair is attempted the compressed air line to the pump should be disconn...

Страница 4: ...Designation Max Solids Passage Inlet Dim Outlet Dim Est Shipping Wt 81 1000 VC4 SS SS FB 76 2 mm 3 76 mm 3 76 mm 3 41 7 kg 92 lbs 81 1002 VC4 SS SS TF 76 2 mm 3 76 mm 3 76 mm 3 41 7 kg 92 lbs 81 1004 VC6 SS SS FB 101 6 mm 4 102 mm 4 102 mm 4 48 9 kg 108 lbs 81 1006 VC6 SS SS TF 101 6 mm 4 102 mm 4 102 mm 4 48 9 kg 108 lbs 81 1008 VC8 SS SS FB 152 4 mm 6 125 mm 6 125 mm 4 69 9 kg 154 lbs 81 1010 VC...

Страница 5: ...ay from the internal vacuum generator s and directed into the vertical column of the pump Once the compressed air is supplied to the top of the process medium the hydraulic force will discharge the pro cess medium from the vertical column These same forces will close the inlet flap valve and open the discharge flap valve simultaneously causing the pro cess medium to move downstream FIGURE 3 The di...

Страница 6: ...381 15 C 157 6 18 N A N A N A N A D 304 12 304 12 304 12 E 101 4 152 6 203 8 F 1486 58 50 1505 59 25 1559 61 38 G 76 3 101 4 102 4 H 76 3 101 4 152 6 J 254 10 254 10 254 10 K 436 17 18 555 21 87 647 25 50 L 352 13 87 352 13 87 352 13 87 S e c t i o n 4 DI M E NSION A L D R A W INGS S A NI F L O T M V C REV A 4 WILDEN PUMP ENGINEERING CO LLC WIL 12120 E 01 ...

Страница 7: ... including pumping distance required flow rate product viscosity and piping size ELECTRICAL The control panel requires 120 220VAC at 1 amps ELEVATION Selecting a site that is well within the pump s dynamic lift capability will assure that loss of prime troubles will be eliminated In addition pump efficiency can be adversely effected if proper attention is not given to site location PIPING Final de...

Страница 8: ... from the pump The Adjustable Probe During operation certain products can coat the sensor probe creating false inputs to the control panel and causing poor performance Periodic cleaning of the sensor probe may be required in certain applications to maintain reliability To decrease the frequency of cleanings the HPVC control panel has an adjustable probe set point This set point can be changed on t...

Страница 9: ... be displayed until the condition is corrected and the pump control is reset This screen alerts the operator to the nature of the problem and provides instructions to solve the problem and restart the pump This screen indicates that product has been in contact with the probe for a period of time that was unexpectedly long To identify the cause the operator should make sure that Air shut off valve ...

Страница 10: ...ow or no air pressure 2 Air valves not functioning properly 3 Dirty or defective vacuum generator 4 Loose connection on piping 5 Check valves misaligned 6 Solution Manually discharge the pump for 2 seconds to clear 1 any blockage If this does not help disassemble the pump and clear the check valves Turn up suction regulator to 5 bar 75 PSI 2 Check to ensure that the air supply is adequate 3 Disass...

Страница 11: ... 2 Begin assembly by installing the flange gasket and inlet flap valve housing to pump base After joining components secure with clamp band NOTE When installing inlet flap valve housing ensure that hinged portion of flap valve is located in the up position by utilizing the UP reference mark on the inlet housing Figure 2 Step 3 Figure 3 Position flange gasket and discharge flap valve housing to pum...

Страница 12: ...ons correctly g Run purge switch functions correctly h Valves function correctly i Pump Cleaning Procedure Empty the discharge piping of product using the 1 purge switch on the control panel Disconnect inlet and outlet piping from the pump 2 keeping clamps and gaskets in a secure place Remove sensor cap and diffuser screen from 3 pump and place them on the sensor cap storage bracket keeping the cl...

Страница 13: ... to the off position and 4 attach a proper lockout device After attaching the lockout devices put the keys in 5 your pocket After locking out the electrical and air supplies try to 6 start the pump to verify that it is properly disabled If the pump starts or still has power applied stop and get your supervisors help Release any stored air in the pump by turning the 7 regulator adjusting knobs full...

Страница 14: ...S e c t i o n 7 E x pl o d e d V i ew P art s L i s t i n g 6 6 6 5 1 2 4 5 3 5 5 SANIFLOTM VC4 E X P L OD E D V I E W 12 WILDEN PUMP ENGINEERING CO LLC WIL 12120 E 01 ...

Страница 15: ...Inlet Check 3 x 3 1 80 0144 80 0144 3 Valve Outlet 4 x 3 Reducer 1 80 0145 80 0145 4 Base 4 Pump 1 80 0084 80 0084 5 Gasket 4 FB Buna 4 N A 80 0101 5 Gasket 4 Teflon 4 80 0152 N A 6 Clamp 4 H D 4 80 0089 80 0089 E x pl o d e d V i ew P art s L i s t i n g SANIFLOTM VC4 part s l i s t i n g 13 WILDEN PUMP ENGINEERING CO LLC WIL 12120 E 01 ...

Страница 16: ...7 5 1 8 8 6 8 6 2 4 3 6 S e c t i o n 8 E x pl o d e d V i ew P art s L i s t i n g SANIFLOTM VC6 E X P L OD E D V I E W 14 WILDEN PUMP ENGINEERING CO LLC WIL 12120 E 01 ...

Страница 17: ...t 6 x 4 Reducer 1 80 0147 80 0147 4 Base 6 Pump 1 80 0085 80 0085 5 Gasket 4 FB Buna 1 N A 80 0101 5 Gasket 4 Teflon 1 80 0152 N A 6 Gasket 6 FB Buna 3 N A 80 0102 6 Gasket 6 Teflon 3 80 0153 N A 7 Clamp 4 HD 1 80 0089 80 0089 8 Clamp 6 HD 3 80 0090 80 0090 E x pl o d e d V i ew P art s L i s t i n g SANIFLOTM VC6 part s l i s t i n g 15 WILDEN PUMP ENGINEERING CO LLC WIL 12120 E 01 ...

Страница 18: ...1 7 5 3 6 8 2 4 6 8 8 6 S e c t i o n 8 E x pl o d e d V i ew P art s L i s t i n g SANIFLOTM VC8 E X P L OD E D V I E W 16 WILDEN PUMP ENGINEERING CO LLC WIL 12120 E 01 ...

Страница 19: ...t 4 Teflon 1 80 0152 N A 6 Gasket 8 FB Buna 3 80 0151 80 0151 6 Gasket 8 Teflon 3 80 0154 80 0154 7 Clamp 4 H D 1 80 0089 80 0089 8 Clamp 8 H D 3 80 0150 80 0150 NOTE The VC8 pump is also available with a 152 mm 6 pump discharge connection To order this other option use item number 81 1011 To order an 8 x 6 outlet reducer for a VC8 pump use item number 81 0149 E x pl o d e d V i ew P art s L i s t...

Страница 20: ...S e c t i o n 8 E x p l o d e d V i e w Pa r t s L i s t i n g vc pump controller E XP L OD E D V I E W 18 WILDEN PUMP ENGINEERING CO LLC WIL 12120 E 01 ...

Страница 21: ...GULATOR GAUGE 28 1 81 12153 8 VACUUM GENERATOR 29 1 81 12286 VACUUM GENERATOR GAUGE 30 1 81 12287 VACUUM GENERATOR MUFFLER 31 1 81 17792 1 4 NPT X 1 4 TUBE SWIVEL 90 32 1 81 12206 SS 1 4 STREET ELL 33 1 81 10938 SS 1 4 CLOSE NIPPLE 34 2 81 11900 SS 1 2 X 1 4 BUSHING 35 3 81 18348 SS 1 2 CLOSE NIPPLE 36 1 81 12497 SS 1 2 CROSS 37 1 81 13883 SS 1 2 BALL VALVE 38 1 81 17779 1 2 NPT X 1 2 TUBE STRAIGH...

Страница 22: ...S e c t i o n 8 v c p u mp c o n tr o ller s c hemat i c 20 WILDEN PUMP ENGINEERING CO LLC WIL 12120 E 01 ...

Страница 23: ...comes first Failure due to normal wear misapplication or abuse is of course excluded from this warranty Since the use of Wilden pumps and parts is beyond our control we cannot guarantee the suitability of any pump or part for a particular application and Wilden Pump and Engineering LLC shall not be liable for any consequential damage or expense arising from the use or misuse of its products on any...

Страница 24: ...nce The result of advanced thought Refine Your Process Designed for sanitary applications Minimize product degradation Improved production yields Easy to inspect clean assemble Minimized water requirements The result of progressive thought Simplify Your Process Maximize Your Process Electronic control monitoring Level control containment Pulsation dampening Drum unloading systems Complete system s...

Отзывы: