48
DC-ITD-FTIR-MAN01
www.dodtec.com
FEB 2020
815-788-5200
FTIR Manual
36
64
Concentration Results During Scan
Real [32] Array with Conc. For Up To
32 gasses
100
4
Scan Date / Time
**
See Note Below
104
64
Alarm Level 1 & 2 Indicators
128 Byte Array
–
Alarm 1&2 for 64
Pts.
168
4
Faults / Alarms FTIR
8 Byte Array Defined Below
172
32
DOD Gas Identification Number
Array of 32 UINT (16 bit) Numbers
204
12
Last Spectra File Name
Array of 26 Characters
**
This is a Microsoft.Net value which represents the number of 100-nanosecond intervals that have elapsed
since 12:00:00 midnight, January 1, 0001, which represents Date, Time, Min. Value. It does not include the
number of ticks that are attributable to leap seconds.
The Fault/Alarms are indicated as follows:
Byte 0 - Critical Fault
Byte 1 - General Fault
Byte 2 - Alarm Level 1
Byte 3 - Alarm Level 2
Byte 4 - Analysis Active
Byte 5
–
Power
Byte 6
–
Testing Mode
Byte 7
–
Heartbeat
F.2 MODBUS TCP/IP
FTIR Faults/Alarms (Discrete Outputs)
Token
Description
Type
Machine Critical Fault
Machine has critical fault
Boolean
Machine Maintenance Fault Machine has a maintenance fault
Boolean
Machine Gas Alarm 1
Level 1 Gas Alarm at Least 1 Machine Point
Boolean
Machine Gas Alarm 2
Level 2 Gas Alarm at Least 1 Machine Point
Boolean
Power On
Power is on to Analyzer
Boolean
Watchdog
Toggles On/Off Every X Seconds
Boolean
Each Point Alarm Indicators
Token
Description
Type
Alarm 1 Active
Level 1 Gas Alarm Active
Boolean
Alarm 2 Active
Level 2 Gas Alarm Active
Boolean
Last Scan
Token
Description
Type
Concentrations
Array of Concentration Values for Last Scan
Real [32]
Sequence#
Sequence Number for Last Scan
Int
Point Valves
Array of Indicators for Active Point(s) during Last Scan Boolean [64]
Scan Date / Time
Date / Time of last Scan
Ticks Since 1/1/1
Gasses Scanned
Array of DOD Gas ID#s In Method
Unit [32]
Spectra Filename
File Name of Spectra File Created
Char [24]
IMPORTANT:
Floating Point numbers on the FTIR are stored Little Endian (least significant register first). The
Master MODBUS system should use "Read Holding Registers” to read data from the FTIR.