DC-ITD-FTIR-MAN01
www.dodtec.com
FEB 2020
815-788-5200
FTIR Manual
Appendix F
–
Optional Communication Interfaces
F.1 Ethernet IP
FTIR Faults/ Alarms (Discrete Outputs)
Token
Description
Type
Machine Critical Fault
Machine has critical fault
Boolean
Machine Maintenance Fault Machine has a maintenance fault
Boolean
Machine Gas Alarm 1
Level 1 Gas Alarm at Least 1 Machine Point
Boolean
Machine Gas Alarm 2
Level 2 Gas Alarm at Least 1 Machine Point
Boolean
Power On
Power is on to Analyzer
Boolean
Each Point Alarm Indicators
Token
Description
Type
Alarm 1 Active
Level 1 Gas Alarm Active
Boolean
Alarm 2 Active
Level 2 Gas Alarm Active
Boolean
Last Scan
Token
Description
Type
Concentrations
Array of Concentration Values for Last Scan
Real [32]
Sequence#
Sequence Number for Last Scan
Int
Point Valves
Array of Indicators for Active Point(s) during Last Scan Boolean [64]
Scan Date / Time
Date / Time of last Scan
Ticks Since 1/1/1
Gasses Scanned
Array of DOD Gas ID#s In Method
Unit [32]
Spectra Filename
File Name of Spectra File Created
Char [24]
Due to limitations in the amount of data available via Ethernet/IP the Originator must track the
concentrations on each point based on the last scan values. To save the concentrations on each point,
monitor for changes in the date/time of the last Scan then update the point concentrations for the point
which has the valve open.
Downlink from Master to FTIR
Assembly Instance :102
Size (16-bit Words): 2
Offset (Words)
Size # Words
Data Description
Format
0
2
Connection Status
Uplink from FTIR to Master
Assembly Instance: 101
Size (16-bit Words): 216
Offset (Words)
Size # Words Data Description
Format
0
2
Connection Status
2
2
Last Scan Sequence Number
UINT (16Bit) Number
4
32
Valves Active During Scan
64 Byte Array (1 Byte Per Valve -
Boolean)