Appendix B
–
I/O Connection Details (Cont.)
B.1 EK1101 Coupler
Each analyzer tray (16 or 32 point) has its own output module section and each section is coupled together
with an EK1101 module. Two different power sources are used for the module. Power to operate the module
itself is supplied by the DOD64 power and prewired by DOD Technologies - Do NOT remove. Power for all
outputs must be supplied by the customer on pins 6 & 7.