![DOD Technologies DOD64-FTIR Скачать руководство пользователя страница 13](http://html1.mh-extra.com/html/dod-technologies/dod64-ftir/dod64-ftir_user-manual_2515355013.webp)
13
DC-ITD-FTIR-MAN01
www.dodtec.com
FEB 2020
815-788-5200
FTIR Manual
Chapter 3
–
Installation
3.1 Selecting A Location
The DOD64-FTIR is designed for safe use under the following conditions:
•
Indoor use only
•
Altitude up to 2,000 m
•
Temperatures 5°C - 40°C
•
Maximum relative humidity 80% for temperatures up to 31°C decreasing linearly to 50% relative
humidity at 40 °C
•
110 VAC or 220 VAC supply voltage fluctuations up to +/- 10% of the nominal voltage
The DOD Technologies Model DOD64-FTIR is supplied with wheels to allow for easy installation. The system
is transported in a wooden crate. The front opening panel door is designed to be used as ramp to roll the
system out of the crate (see installation package for details). The DOD64-FTIR rests on four rollers which can
be locked in place and bolted to the floor if necessary.
CAUTION:
Care must be taken when loading, unloading and moving the DOD64-FTIR. The DOD64-FTIR is a
heavy piece of equipment which could cause injury or death if not handled properly. Make sure the rollers are
operating properly and only move the DOD64-FTIR on a level surface.
The DOD64-FTIR should be placed in a location as central as possible to the locations being monitored while
considering the following restrictions:
•
The maximum sample line length is 400 ft. (121 m). Using the shortest possible sample line length will
reduce transport times and increase the response time of the DOD64-FTIR. (see Appendix C)