4:
Troubleshooting
69
Illumination button is not activating
Possible Reason
Action
Loose internal power supply button cables
Please contact your distributor
Button is receiving incorrect voltage
Please contact your distributor
Dimmer button is not activating
Possible Reason
Action
Loose internal power supply button cables
Please contact your distributor
Button is receiving incorrect voltage
Please contact your distributor
Improper illumination intensity
Possible Reason
Action
Dimmer dial is set too low, so there is not enough
light
Use the dimmer dial to increase the illumination
Button is receiving incorrect voltage
Please contact your distributor
Low incoming LED voltage
Please contact your distributor
Faulty LED
Please contact your distributor
Dark box light penetration
Possible Reason
Action
Improper positioning of the drawer
Verify that the drawer is inserted properly
Содержание MicroChemi
Страница 1: ...Welcome to the MicroChemi System...
Страница 16: ...3 Setup and Installation 15 4 Mark the checkbox if you wish to create a desktop icon and click Next...
Страница 20: ...3 Setup and Installation 19 6 Click the combo box and select the MicroChemi component from the list...
Страница 34: ...3 Setup and Installation 33 14 Click OK 15 Choose DNR Camera and click Next...