12
7117100018L30 1712V002
6.5 Information about electrical
connections
i
Ensure that electrical connections to the
mains power supply are carried out in accord-
ance with current valid national and local reg-
ulations and standards governing the installa-
tion of low voltage units in medical facilities�
i
Install an all-pole disconnect switch with a
contact opening width of at least 3 mm in
the electrical connection to the mains power
supply�
i
Observe the current consumption of the
devices that are to be connected�
i
Install electrical lines without mechanical ten-
sion�
i
Make the electrical connection via the main
power switch of the treatment unit or via the
main power switch of the practice�
6.6 Information about connecting
cables
Mains supply cable
Installation type
Line layout (minimum
requirements)
Fixed installation
– Plastic sheathed cable
(e�g� type NYM-J)
Flexible
– PVC flexible line
(e�g� H05 VV-F)
or
– Rubber connection
(e�g� H05 RN-F or
H05 RR-F)
6 Requirements
6.1 Installation/setup room
The room chosen for set up must fulfil the fol-
lowing requirements:
– Closed, dry, well-ventilated room
– Should not be a room made for another pur-
pose (e� g� boiler room or wet cell)
6.2 Setup options
CAS 1 Combi-Separator
– Directly in the treatment unit�
– In a special housing in an extension of the
treatment unit�
6.3 Hose materials
For waste connections and suction lines only
use the following hose types:
– Flexible spiral hoses made of PVC with inte-
grated spiral or equivalent hoses
– Hoses that are resistant to dental disinfectants
and chemicals
Plastic hoses will display signs of ageing
over time� Therefore, they should be
inspected regularly and replaced as nec-
essary�
The following types of hoses must not be
used:
– Rubber hoses
– Completely PVC hoses
– Hoses that are not sufficiently flexible
6.4 Installation and routeing of hos-
es and pipes
i
Execute the on-site pipe installation in accord-
ance with the applicable local regulations and
standards�
i
Lay the hose installation of the drains to or
from the unit at a sufficient incline�
If incorrectly laid, the hoses can
become blocked with sedimentation�
Installation
EN
Содержание L2-S300
Страница 41: ...41 Cleaning and Disinfection of the Suction Handpieces...
Страница 42: ...42 Cleaning and Disinfection of the Suction Handpieces...
Страница 66: ...Instructions for use Electric motor EM 12 L Supply hose VE 10...
Страница 79: ...3 Product description Motor EM 12 L Motor shealt LED Connection for instruments as per ISO 3964 O rings Seal 14...
Страница 80: ...Supply hose VE 10 Connection sheat Connection quick release coupling 15...
Страница 110: ......
Страница 112: ...Instructions for use Handpiece PB 5 L PB 5 L S PB 5 L Q...
Страница 126: ...15 3 Product description Thread Connection for supply hose Handpiece cap Optical fibre Optic outlet Nozzle cleaner...
Страница 157: ......
Страница 160: ...Instructions for Use Foot control S NW S N2 S N1...
Страница 191: ...32 Manufacturer s declaration...
Страница 192: ...33 Manufacturer s declaration...
Страница 193: ...34 Manufacturer s declaration...
Страница 194: ...35 Manufacturer s declaration...
Страница 196: ...1 Minilight Installation and use manual...
Страница 207: ...12 14 WIRING DIAGRAM 14 1 VERSION 6F 14 2 VERSION L...
Страница 222: ...EN 1801V002 7560100003L02 Installation and Operating Instructions Spittoon valve 3...
Страница 223: ......
Страница 246: ......
Страница 247: ......
Страница 248: ......
Страница 250: ...EN 1712V002 7117100018L30 Installation and Operating Instructions CAS 1 Combi Separator...
Страница 251: ......
Страница 268: ...7117100018L30 1712V002 17 Installation 9 Service program EN...
Страница 284: ......
Страница 286: ...EN 1709V004 9000 606 39 30 Installation and Operating Instructions CS 1 Combi Sepamatic...
Страница 287: ......
Страница 309: ......
Страница 310: ......
Страница 311: ......
Страница 312: ......
Страница 314: ......
Страница 327: ...EVA USER MANUAL EN Edition 1 0 April 2020 Pag 14 di 14...