4
Important Information
This manual complies with the standards and directives mentioned in the EN
DIN 61010-1 Norm. IQ, OQ and PQ documents are available on request.
The systems are provided with a high-precision self-optimising
temperature controller. Due to the self-optimisation it might oc-
cur that the set temperature is exceeded in individual cases.
Please note, that by heating your samples dangerous gases may
be emitted. In this case, the device must be used with an extrac-
tor hood.
If the ThermoMixers are operated with a high shaking frequency,
the vibrations of the system might transmit to the underground.
Please do consider this when placing the system.
Contact with highly flammable liquids must be avoided. The ex-
changeable block may be hot.
The ventilation of the device must not be blocked. To ensure the
proper ventilation at all times please consider a minimum
distance of 10 centimeter on each side of the device.
Environmental conditions
The ideal area of operation of the devices is achieved at 80% relative air
humidity at most and lays b3°C - +50°C ambient air temperature.
This values should not be under run or exceeded.
i
Содержание HLC MHL 23
Страница 25: ...25 800 68 13 02_Manual_Smart_Control docx...
Страница 26: ...26...