background image

 

Important Information 
 

This manual complies with the standards and directives mentioned in the EN 
DIN 61010-1 Norm. IQ, OQ and PQ documents are available on request. 

 
 
The systems are provided with a high-precision self-optimising 
temperature controller. Due to the self-optimisation it might oc-
cur that the set temperature is exceeded in individual cases.  
 
Please note, that by heating your samples dangerous gases may 
be emitted. In this case, the device must be used with an extrac-
tor hood.  
 
If the ThermoMixers are operated with a high shaking frequency, 
the vibrations of the system might transmit to the underground. 

        Please do consider this when placing the system. 

 

Contact with highly flammable liquids must be avoided. The ex-

changeable block may be hot. 
 

The ventilation of the device must not be blocked. To ensure the 

proper ventilation at all times please consider a minimum  

distance of 10 centimeter on each side of the device. 

 

Environmental conditions

 

 

The ideal area of operation of the devices is achieved at 80% relative air  

humidity  at most and  lays  b3°C -  +50°C  ambient air temperature. 

This values should not be under run or exceeded. 

 

 

Содержание HLC MHL 23

Страница 1: ...1 Thermo Shaker and BlockThermostats with Smart Control MKR 13 MKR 23 MHR 13 MHR 23 MHL 23 TK 23 Operating Manual...

Страница 2: ...4 Temperature accuracy 5 First Steps 5 Setup 5 Operation 6 Parameter Setting 6 Working with the Programming Function 10 Assembly of Accessories 13 Anti Condensation Plate BA 24 96 800012900 800013000...

Страница 3: ...s must be avoided The ventilation of the device must not be blocked To ensure the proper ventilation at all times please consider a mini mum distance of 10 centimeter on each side of the device Damage...

Страница 4: ...extrac tor hood If the ThermoMixers are operated with a high shaking frequency the vibrations of the system might transmit to the underground Please do consider this when placing the system Contact wi...

Страница 5: ...red When setting a process time please note that the sample temperature is always reached a little later than the displayed block temperature First Steps Setup Plug the mains cable in the IEC socket o...

Страница 6: ...stem Setup menu Push the up arrow key for 5 seconds then the setup menu is displayed With and the values can be selected for setting The selected box flashes with ok it will be permanently set for cha...

Страница 7: ...hange can also be made in the same way during a running process After confirming the input with ok the new set value is effective immediately For the hundreds the following values can be selected 0 fo...

Страница 8: ...endless process select 00 00 100 Short Mix vortexing Navigate with and to the button 100 Short Mix and confirm with ok The system shakes with maximum mixing speed regardless of a process is started or...

Страница 9: ...emaining time is displayed in the menu The time only starts to run if the temperature set value is reached This does also apply for subsequent tempera ture changes A set value change can also be made...

Страница 10: ...layed after the expiration of the time The device turns off automatically Working with the Programming Function Program menu Navigate with and to the button Pro grams and confirm with ok A total of 30...

Страница 11: ...t ting is only required if the sample is to be tempered slower than technically possi ble If the sample is to be tempered as fast as possible set the value to 0 Rotational speed Shaking frequency with...

Страница 12: ...ng program The set time is the time of the running step the remaining time is the time of the complete process In the lower right part of the display program no cycle and step are indicated Terminatio...

Страница 13: ...hold of the bolt of the system and with that tightly attaches block and anti condensation to the system Data Transfer A USB 1 1 interface for the communication with a PC is provided as a stand ard Con...

Страница 14: ...hh mm tmp rpm prg cycle step steptime hh mm rem risetime_m mix_s pause_m 00 00 26 9 0000 01 01 01 00 00 00 9 0 2 00 01 33 7 0220 01 01 01 00 00 00 9 0 2 00 02 36 7 0000 01 01 01 00 00 00 9 0 2 00 03 3...

Страница 15: ...the next valid value r 020 Changes the set rotational speed to a new value The rotational speed must always be entered with three digits r 120 sets 1 200 1 min r 045 sets 450 1 min r is the identific...

Страница 16: ...correct system type name at power on If not please contact the HLC by DITABIS service department There are excessive temperature fluctuations Check the seat of the changeable block by pulling it up If...

Страница 17: ...slit Do not perform spray disinfection Do only connect the system with the power supply if it is completely dry The repair service may only be performed by staff authorized and trained by the manufac...

Страница 18: ...off all outer parts of the system with the disinfectant and a lint free cloth In Case of Service If a technical problem arises please contact your local dealer or the DITABIS service department The D...

Страница 19: ...ity 1 changeable block Weight 9 kg Electr heating cooling power 130 W Electr supply 230 V 50 Hz Technical Data MKR 23 Temperature working range Room temp 11 C to 70 C Temperature adjustable range 0 C...

Страница 20: ...y 1 changeable block Weight 7 kg Electr heating cooling power 200 W Electr supply 230 V 50 Hz Technical Data MHR 23 Temperature working range Room temp 3 C to 130 C Temperature adjustable range 0 C to...

Страница 21: ...ity 2 changeable blocks Weight 7 kg Electr heating cooling power 350 W Electr supply 230 V 50 Hz Technical Data TK 23 Temperature working range Room temp 16 C to 90 C Temperature adjustable range 10 C...

Страница 22: ...deep flat bottom 800011900 BP 28 for 8 x D 28 mm 40 mm deep flat bottom For centri 800012200 BZ 15 Falcon tubes 14 x 15 ml with isolation lid fuge tubes 800012300 BZ 50 for Falcon tubes 6 x 50 ml wit...

Страница 23: ...T 01 Tempering tub for 1 block transparent 800013900 BT 02 Tempering tub for 2 blocks transparent 800012900 BA 24 Anti condensation plate for 1 block micro test tubes 800013000 BA 96 Anti condensation...

Страница 24: ...24 Explanations Caution Risk of electric shock Caution Caution Hot surface Important information Earth conductor Fuse i...

Страница 25: ...25 800 68 13 02_Manual_Smart_Control docx...

Страница 26: ...26...

Отзывы: