9
Process without time setting
Only set a set temperature and a set shaking
frequency but set the time on continuous
00:00
.
Now, the process time is displayed
continuously.
It only starts to run if the temperature set
value is reached.
A set value change can also be made as de-
scribed above during a running process.
Changes are retained also after the comple-
tion of the process and are displayed in the
start menu as current set values.
For terminating the process, push
Stop
.
Process with time setting
If a process time is set, the remaining time is
displayed in the menu. The time only starts to
run if the temperature set value is reached.
This does also apply for subsequent tempera-
ture changes. A set value change can also be
made as described above during a running
process.
The process ends with or without a signal,
depending on the settings in the setup menu.
For changes during the process
If set value changes have been made, you can
select to save these changes and to have
them displayed in the start menu as new set
values.
Stopping the process before the expiration
of the time
If the process is stopped before the expiration
of the time, you can select whether the pro-
cess really should be terminated or i fit
should continue. Depending on the selection,