Anti-Condensation Plate BA 24 / 96 (800012900 / 800013000)
The anti-condensation plate BA 96 (800013000) is placed on the attached
block, no assembly steps are required.
For using the anti-condensation plate BA 24 (800012900), remove the socket
screw in the block with the enclosed screwdriver. Screw the enclosed thread-
ed pin with isolation knob into the anti-condensation plate. Place the anti-
condensation plate on the block and screw the threaded pin in deeper so that
it takes hold of the bolt of the system and with that tightly attaches block and
anti-condensation to the system.
Data Transfer
A USB 1.1. interface for the communication with a PC is provided as a stand-
ard. Connect the system via the USB-port on the left side with a USB-cable to
your PC. If this is not available at your computer, install the suitable USB-
driver. You will find it under
www.ftdichip.com/Drivers/VCP.htm
). These USB-
drivers generate a new virtual COM-interface in your PC. The chip in the Smart
Control is called FT232B.
Access process data via USB
Start a terminal program (e.g. Hyper terminal) on your PC.
This program you will find under Windows in "Programs" / "Accessories" /
"Communication". The window "Hyper Terminal" is opened. Proceed system-
atically, please note that for the connection setting for "Bits per second"
115200 is selected.