15
Starts a process without program with the parameters indicated on the dis-
play.
Time settings, if applicable, are deleted and overwritten by continuous since
time settings must be programmed on the computer.
Stop
Stops all processes and returns to the start menu, running programs / pro-
cesses are aborted.
t=0370
Changes the set temperature to a new value.
The temperature must always be entered with four digits.
t=0370 sets 37.0°C, t=0050 sets 5.0°C, t=-060 sets -6.0°C.
t= is the identification for a temperature value. The associated numerical
value reflects the value in tenth of a degree.
The first digit after the = can be a 0, a 1 or a hyphen. Invalid temperature
values are - like for the input at the system - replaced by the next valid value.
r=020
Changes the set rotational speed to a new value.
The rotational speed must always be entered with three digits.
r=120 sets 1,200 1/min, r=045 sets 450 1/min.
r= is the identification for a rotational speed value. The associated numerical
value reflects the value in 10 1/min.
Invalid rotational speed values are - like for the input at the system - replaced
by the next valid value.
r=000
Stops the rotational speed, e.g. if shaking should be stopped but tempering be
continued.
Other inputs like e.g. r=0 are interpreted as rotational speed and cause the
motor to rotate with the minimum rotational speed.
Every command must be confirmed with "CR"
In case of unknown commands and/or in case of spelling errors, nothing is
done.
The inputs in the setup menu like colour, contrast and brightness of the dis-
play, language, rotation direction etc. can only be done at the operating panel
of the system itself.
Содержание HLC MHL 23