25
1-800-245-9917
Event Counter
Your Attendant Bluetooth Digital Chair Scale has been designed with an Event Counter type of security seal.
When selected, the scale will display two numbers representing the Calibration and Configuration counters.
Calibration Counter (CAL)
The calibration counter is incremented when a value in the calibration part of setup is changed (
CAP
,
int
,
dEC
,
Sr
,
Unit
,
FiLt
,
nnOt
,
EURO
,
CAL
,
GrAu
,
TYPE
). The counter is only incremented one time even if
more than one parameter is changed each time through setup.
Configuration Counter (CFg)
The configuration counter is incremented when a value in the configuration part of setup is changed (
PUO
,
Otr
,
SEr1
,
SEr2
,
USb
,
YEAr
,
nntH
,
dAY
,
HOUr
,
nnIn
,
SEC
,
S Ht
,
Hold
). The counter is only incremented
one time even if more than one parameter is changed each time through setup.
To Review the Event Counter:
1. Press the key to turn the scale on.
2. The scale will perform a display test (turn on all segments and annunciators) and then change to the
show the software revision for a few seconds.
3. Next, the display will change to the Weight Display with the
0
(ZERO) and
(STABLE)
annunciators turned on to show a stable, center-of-zero weight condition.
4. Press and hold the
key.
5. The display will change to show all dashes and then the software revision.
6. Release the
key.
7. The display will change to show
CAL
(Calibration Counter) on the small upper seven-segment display
and up to a 3-digit number on the large lower seven-segment display for approximately two (2)
seconds.
8. Next, the display will then show
CFg
(Configuration Counter) on the small upper seven-segment
display and up to a 3-digit numbers on the large lower seven-segment display for approximately two (2)
seconds and then change to show all dashes.
•
NOTE:
If your scale is model 6880DS, additional information for Wi-Fi and Bluetooth (BLE) settings will be
shown after the Configuration Counter. Refer to the Bluetooth Low Energy (BLE) section for instructions.
9. To return to the normal operation, press the
key.
10. Otherwise, press the key to turn the scale off.