21
1-800-245-9917
(SECONDS)
The display will show
SEC
and the current setting. If the value displayed is acceptable,
press the key. Otherwise, press the or keys to select a new value and then press
the key to save it and proceed to the next prompt.
This is the current seconds setting of the real-time clock. Allowable values are 0 to 59.
•
NOTE:
The following two parameters are not used with the Model 6880DS scale. Press
the key to save the setting displayed and proceed to the next prompt.
(Sonar Height Calibration)
The display will show
HtCL
and the current setting
no
. Press the key to save the setting
and proceed to the next prompt.
(SENSOR HEIGHT)
The display will show
S Ht
and the current setting. Press the key to save the setting and
proceed to the next prompt.
(HOLD TIME)
This setting is used by the scale to hold a stable user weight for a desired amount of time.
For example, if a value of 5 seconds is used, then when the scale locks onto a stable user
weight, it will remain locked for 5 seconds before automatically releasing the weight.
The display will show
Hold
and the current setting. If the value displayed is acceptable,
press the key. Otherwise, press the or keys to select a new value and then press
the key to save it and proceed to the next prompt.
This setting controls the auto-locking feature. Allowable values for hold time are 0 to 20.
(CALIBRATION)
The display will show
CAL
and the current setting
no
. If the scale has been previously
calibrated and you wish to skip calibration and proceed to the
tYPE
(Scale Model) prompt,
press the key and the previous calibration will be retained.
To begin calibration, press the key to select
YES
and then press the key. After
pressing the key, the display will change to
LOAd
.