18
Setup & Calibration
(cont.)
(MOTION RANGE)
The display will show
nnOt
and the current setting. If the value displayed is acceptable,
press the key. Otherwise, press the or keys to select a new value and then press
the key to save it and proceed to the next prompt.
The motion range is the number of divisions of change permitted before indicating unstable
(the STABLE annunciator turns off). Allowable values are 1 to 10.
(OIML)
The display will show
EUrO
and the current setting. If the value displayed is acceptable,
press the key. Otherwise, press the or keys to toggle the OIML selection between
YES
or
no
and then press the key to save it and proceed to the next prompt.
This setting controls the OIML (European) specific requirements. Allowable settings are
YES
for EU or
no
for US.
(Power Up Zero)
The display will show
PUO
and the current setting. If the value displayed is acceptable,
press the key. Otherwise, press the or keys to toggle the power up zero between
YES
or
no
and then press the key to save it and proceed to the next prompt.
This setting controls zeroing of the scale on power up. If enabled, the scale will be reset to
zero automatically on power up. Allowable settings are
YES
or
no
.
(Zero Tracking)
The display will show
Otr
and the current setting. If the value displayed is acceptable, press
the key. Otherwise, press the or keys to select a new value and then press the
key to save it and proceed to the next prompt.
This setting controls the automatic zero tracking of the scale. Allowable values are 0 to 10,
indicating the number of half-divisions above or below zero that the scale will attempt to
maintain the zero position.
NOTE:
Select 0 (zero) to disable zero tracking.
(Com1 Mode)
The display will show
SEr1
and the current setting. If the value displayed is acceptable,
press the key. Otherwise, press the or keys to toggle the Com1 Mode between
Cont
(continuous output),
POLL
(weight on demand) and
SPot
(Welch-Allyn Spot LXi
interface) then press the key to save it and proceed to the next prompt.
This setting controls the output for Com 1. Allowable settings are
Cont
,
POLL
or
SPot
.
If
Cont
(continuous output) was selected for Com 1, the scale will continuously transmit
weight data.
If
POLL
(Weight-On-Demand) was selected for Com 1, and the scale is connected to a
computer for transmission of weight data to a PC-based EMR (electronic medical record)
software program, it will transmit a single set of weight data each time the computer sends
a weight request “ENQ” (hex 05) or a SMA weight request “W”.
If
SPot
(Welch-Allyn Spot LXi interface) was selected for Com 1, the scale will automatically
interface to a Welch-Allyn Spot LXi interface device to transmit weight data.