Figure 2-22. AS40 Autosampler Connection
ICS-90 Installation Instructions
36
Doc. 031856-03 12/06
c.
Plug the 2-pin connector from the pair of wires labeled
LOAD
on the
Relay Control cable into the
TTL Output
connector on the ICS-90.
2.17 Operating the ICS-90
ICS-90 installation is complete. For routine operation, refer to the
ICS-90 Ion
Chromatography System Operator’s Manual
(Document No. 031851).
2.18 Installation Troubleshooting
Problem:
When the ICS-90 power is turned on for the first time, a Windows
message box appears asking for a USB configuration file (cmwdmusb.inf).
Possible Cause:
The USB cable was connected and the power turned on before
Chromeleon or Chromeleon Xpress was installed.
Solution:
1. Click the
Cancel
button in the Windows message box.
2. Turn off the ICS-90 power and unplug the USB cable from the computer.
3. Install Chromeleon or Chromeleon Xpress (see
4. Reconnect the USB cable to the computer and turn on the ICS-90 power.
Windows will now automatically recognize the new ICS-90.